{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Scarborough",
      "state": "ME",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98400",
      "recalling_firm": "Abbott Diagnostics Scarborough, Inc.",
      "address_1": "10 Southgate Rd",
      "address_2": "N/A",
      "postal_code": "04074-8303",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1565-2026",
      "product_description": "Product Name: ID NOW\" Influenza A & B 2  Model/Catalog Number: 427-000  Software Version: Not Applicable  Product Description: Influenza A/B virus nucleic acid IVD, kit, nucleic acid technique  Component: Not Applicable",
      "product_quantity": "111984 units",
      "reason_for_recall": "It was confirmed that the impacted lot has a higher occurrence of invalid rates when compared to the product Instructions for Use.",
      "recall_initiation_date": "20260209",
      "center_classification_date": "20260313",
      "report_date": "20260325",
      "code_info": "List Number: 427-000; UDI-DI: 10811877010422; Lot Number: 000X133126; Expiry: 2027 02 21;",
      "more_code_info": ""
    }
  ]
}