  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78412",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and in the countries of  Canada, Puerto Rico, United Arab Emirates, Philippines, and  Europe.",
      "recall_number": "Z-1562-2018",
      "product_description": "AirLife Resuscitation, Adult, with mask, 28inch large bore tubing    Product Usage:  The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.",
      "product_quantity": "8,100 units",
      "reason_for_recall": "The mask component on various lots of the resuscitation devices have been identified as having the  potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the  mask from the elbow of the resuscitator.",
      "recall_initiation_date": "20170427",
      "center_classification_date": "20180510",
      "termination_date": "20190509",
      "report_date": "20180516",
      "code_info": "Part Number: 2K8017. Lots: 0001022516, 0001025656, 0001025655, 0001029028, 0001040668"
    }
  ]
}