{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Fort Lauderdale",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98317",
      "recalling_firm": "Trividia Health, Inc.",
      "address_1": "2400 Nw 55th Ct",
      "address_2": "N/A",
      "postal_code": "33309-2672",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.",
      "recall_number": "Z-1546-2026",
      "product_description": "Owner's Booklets and Instructions for Use that are used with the following blood glucose measurement devices.  Description/Item:  TRUE METRIX PRO Meter Only/RE4H01P-40,  TRUE METRIX PRO Starter Kit/RE4H01P-43,  PROCURE TRUE METRIX PRO Meter Only/RE4212P-00,  Henry Schein TRUE METRIX PRO Meter Only/RE4099P-40,  Henry Schein TRUE METRIX PRO Kit/RE4099P-43,  McKESSON  TRUE METRIX PRO Meter Only/RE4051P-00,  Moore Medical TRUE METRIX PRO Meter Only/RE4068P-01.",
      "product_quantity": "1,001,534",
      "reason_for_recall": "The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.",
      "recall_initiation_date": "20260206",
      "center_classification_date": "20260312",
      "report_date": "20260318",
      "code_info": "Device Item/UDI-DI:   RE4H01P-40/21292006099, RE4H01P-43/21292006105, RE4212P-00/10840330705681, RE4099P-40/21292014919, RE4099P-43/21292008673, RE4051P-00/20612479197217, RE4068P-01/607415261097.  Owner's Booklet or IFU Description/Part and Revision:  TRUE METRIX PRO Owners Booklet/RE4TVHP03 Rev 56 and prior, McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior,   Device serial numbers: Prefix \"TP\" followed by a seven-digit numeric sequence."
    }
  ]
}