  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87659",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St Bld 950",
      "address_2": "",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of FL, MA, MD, MI, MO, NC, NJ, NY, PA, PR, SC, TN, TX.",
      "recall_number": "Z-1536-2021",
      "product_description": "AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies.  Fresenius P/N 24-2500-0    All 3 units make up the AquaBplus 2500Fresenius Part Number:  G02040109-US AquaBplus 3000 - Typ USA  G02040115-US AquaBplus 2500 - Typ USA  G02040120FMC-US AquaBplus HF- Typ USA",
      "product_quantity": "27 units",
      "reason_for_recall": "AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation",
      "recall_initiation_date": "20210318",
      "center_classification_date": "20210507",
      "termination_date": "20240410",
      "report_date": "20210519",
      "code_info": "Serial number first stage (AquaBplus):  5BPS0601  6BPS0933  7BPS1457  7BPS1526  7BPS1458  7BPS1527  8BPS1801  8BPS1800  8BPS1870  8BPS1883  8BPS1898  8BPS1885  8BPS1884  8BPS1918  7BPS14443  8BPS1930  8BPS1928  8BPS1929  8BPS1937  8BPS1927  8BPS2052  8BPS2081  8BPS2108  8BPS2157  9BPS2234  7BPS1441  9BPS2512.    Serial number second ¿stage(B2):  5BDS0194  6BDS067  7BDS0692  7BDS0726  7BDS0643  7BDS0722  8BDS0907  8BDS0909  8BDS0958  8BDS0960  8BDS0964  8BDS0963  8BDS0965  8BDS0978  7BDS0684  8BDS0989  8BDS0997  8BDS0995  8BDS0998  8BDS0992  8BDS1075  8BDS1087  8BDS1102  8BDS1139  9BDS1181  7BDS0683  9BDS1373.    Serial number HF option:  5BFS0125  6BFS0241  7BFS0413  7BFS0456  7BFS0446  7BFS0486  8BFS0541  8BFS0609  8BFS0644  8BFS0665  8BFS0663  8BFS0577  8BFS0621  8BFS0616  7BFS0439  8BFS0686  8BFS0702  8BFS0703  8BFS0695  8BFS0676  8BFS0727  8BFS0757  8BFS0776  8BFS0805  9BFS0845  7BFS0477  9BFS0997",
      "more_code_info": ""
    }
  ]
}