  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Carmel",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94302",
      "recalling_firm": "SonarMed Inc",
      "address_1": "12220 N Meridian St Ste 150",
      "address_2": "",
      "postal_code": "46032-6972",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "_AK, AZ, CA, FL, IL, KY, MD, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, & WA_",
      "recall_number": "Z-1535-2024",
      "product_description": "AW-M0001 SONARMED MONITOR. Used in conjunction with SonarMed Acoustic Sensors for airway monitoring.",
      "product_quantity": "145 units",
      "reason_for_recall": "Software anomaly resulted in failure to detect a partial obstruction in 2.5 mm sensors and up to 3mm distal to the sensor.",
      "recall_initiation_date": "20240325",
      "center_classification_date": "20240418",
      "report_date": "20240424",
      "code_info": "Model No. AW-M0001; GTIN: 00851334007001; Serial No. AW0572, AW0573, AW0574, AW0577, AW0578, AW0579, AW0580, AW0581, AW0584, AW0585, AW0586, AW0588, AW0591, AW0593, AW0594, AW0600, AW0606, AW0607, AW0615, AW0616, AW0624, AW0626, AW0627, AW0648, AW0650, AW0651, AW0658, AW0664, AW0672, AW0674, AW0675, AW0678, AW0680, AW0681, AW0684, AW0685, AW0701, AW0702, AW0703, AW0704, AW0705, AW0706, AW0707, AW0708, AW0709, AW0710, AW0711, AW0712, AW0718, AW0719, AW0720, AW0721, AW0722, AW0723, AW0724, AW0725, AW0726, AW0727, AW0728, AW0729, AW0730, AW0731, AW0732, AW0733, AW0734, AW0735, AW0736, AW0737, AW0738, AW0739, AW0740, AW0741, AW0742, AW0743, AW0744, AW0745, AW0746, AW0747, AW0748, AW0749, AW0763, AW0764, AW0765, AW0766, AW0767, AW0768, AW0769, AW0770, AW0771, AW0772, AW0773, AW0774, AW0775, AW0776, AW0777, AW0778, AW0779, AW0780, AW0781, AW0782, AW0783, AW0784, AW0785, AW0786, AW0787, AW0788, AW0789, AW0790, AW0791, AW0792, AW0793, AW0794, AW0795, AW0796, AW0797, AW0798, AW0799, AW0800, AW0804, AW0805, AW0814, AW0817, AW0818, AW0819, AW0820, AW0821, AW0825, AW0826, AW0827, AW0828, AW0830, AW0831, AW0832, AW0833, AW0834, AW0835, AW0836, AW0837, AW0838, AW0839, AW0840, AW0841, AW0842, AW0854, AW0855.",
      "more_code_info": ""
    }
  ]
}