  "results": [
    {
      "status": "Terminated",
      "city": "Huntingdon Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79872",
      "recalling_firm": "Menarini Silicon Biosystems",
      "address_1": "3401 Masons Mill Rd Ste 100",
      "address_2": "N/A",
      "postal_code": "19006-3574",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of :  Canada and Italy",
      "recall_number": "Z-1530-2018",
      "product_description": "CELLSEARCH¿ CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001)    CELLSEARCH¿ CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.",
      "product_quantity": "N/A",
      "reason_for_recall": "Unusually high number of total images/unassigned events may require additional time and effort for the operator to review all the images and may cause a delay in reporting the results.",
      "recall_initiation_date": "20180404",
      "center_classification_date": "20180426",
      "termination_date": "20190409",
      "report_date": "20180502",
      "code_info": "UDI  (01)10888385000012(17)180810(10)S038  (01)10888385000012(17)180810(10)  (01)10888385000012(17)180810(10)S038S  (01)10888385000012(17)180810(10)S038T  (01)10888385000012(17)180810(10)S038U  (01)10888385000012(17)180810(10)S038 V    Serial, Batch, Lot #  2038  S038R  S038S  S038T  S038U  S038V  S038W  (0 distributed to the field)"
    }
  ]
}