  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Yokneam",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70863",
      "recalling_firm": "Lumenis Limited",
      "address_1": "13 Hayetzira St.,Yokneam Ind. Park",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide (to OR, NM, NY, IL, and CO), Switzerland, Canada, Italy, Great Britain, United Arab Emirates,  France, Belgium, India, Israel, Japan, Portugal, Czech Republic, Kuwait, Germany, Spain, Georgia, Armenia, Slovenia, New Zealand, Saudi Arabia, and Russian Federation.",
      "recall_number": "Z-1519-2015",
      "product_description": "Light Sheer Desire Diode Laser System with XC Handpiece Accessory options.",
      "product_quantity": "117 units",
      "reason_for_recall": "Device software treatment preset parameters for the XC treatment handpiece do not match the Operator Manual, and exceed recommended settings. Operator Manual parameters are lower than indicated for specific hair color and Fitzpatrick skin type resulting in insufficient treatment effect.  May result in patient burns and hypopigmentation.",
      "recall_initiation_date": "20150325",
      "center_classification_date": "20150424",
      "termination_date": "20160615",
      "report_date": "20150506",
      "code_info": "Lot numbers: XI0017, XI0018, XI0023, XIP0043, 182, 184, 185, XIP0049, 236, 237, 239, 240, 242, 248, 249, 250, 251, 254, 256, 257, XIP0038, XIP0131, XIP0125, 3048, 3052, 3065, 3074, XI0012, XIP0143, XIP0144, XI0001, XIP0072, XI0028, XIP0061, XI0028, XI0036, XIP0074, XIP0077, XIP0082, XIP0087, XIP0132, XIP0127, XIP0124, XIP0129, XI0008, XI0029, XI0037, XI0046, XIP0123, XIP0059, XIP0107, XIP0089, XIP0096, XIP0156, XI0005, XIP0114, XI0014, XI0022, XI0015, XI0021, XI0031, XI0034, XI0024, XI0026, XIP0047, XI0042, XI0039, XIP0044, XIP0025, XIP0035, XIP0041, XIP0030, XIP0051, XIP0057, XIP0113, XIP0067, XIP0053, XIP0083, XIP0058, XIP0055, XIP0060, XIP0066, XIP0163, XIP0071, XIP0063, XIP0120, XIP0088, XIP0081, XIP0064, XIP0050, XIP0101, XIP0062, XIP0095, XIP0121, XIP0094, XIP0108, XIP0105, XIP0122, XIP0119, XIP0104, XIP0097, 104, XIP0065, XIP0164, XIP0159, XIP0152, XIP0166, XI0016, XIP0045, XIP0052, XIP0158, XIP0092, XIP0111, XIP0086, XIP0086, XIP0086, XIP0112, XIP0130, XIP0157, XIP0093, XIP0151, XIP0149, XIP0167, XIP0168, XIP0162, XIP0171, XIP0138, XIP0118",
      "more_code_info": ""
    }
  ]
}