  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84947",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, MI, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI. SC, TN, TX, UT, VA, WA, WI, WY, and countries of AE, AF, AL, AO, AR, AT, AU, AZ, BD, BE, BO, BG, BH, BR, CA, CH, CO, CL, CN, CO, CR, CY, DE, DO, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, IE, IN, IR, IS, IT, IQ, JO, KE, KR, LB, LT, LV, LY, ME, MK, MX, NL, NO, PA, PE, PH, PK, PL, PS, PT, PY, QA, RO, RU, SA, SE, SK, SL, SV, SY, TH, TJ, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XK, XS, ZA",
      "recall_number": "Z-1514-2020",
      "product_description": "IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS: L2KE22 (200 tests), L2KE26 (600 tests) - Product Usage: For in vitro diagnostic use with the IMMULITE systems Analyzers   for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various sexual disorders.",
      "product_quantity": "6118 kits",
      "reason_for_recall": "Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.",
      "recall_initiation_date": "20200205",
      "center_classification_date": "20200318",
      "termination_date": "20211118",
      "report_date": "20200325",
      "code_info": "Manufacturing Lot #: All kit lots beginning 501 and above. SMN & (UDI) for L2KE22(D) US: 10702833 (00630414947907), L2KE26(D) 10702834 (00630414947914). SMN & (UDI) for L2KE22 OUS: 10381178 (00630414966106), L2KE26: 10381177 (00630414966007).",
      "more_code_info": ""
    }
  ]
}