  "results": [
    {
      "status": "Ongoing",
      "city": "Snoqualmie",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91851",
      "recalling_firm": "Spacelabs Healthcare, Inc.",
      "address_1": "35301 Se Center St",
      "address_2": "",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.",
      "recall_number": "Z-1503-2023",
      "product_description": "Ultraview SL (UVSL) Command Module, Model 91496",
      "product_quantity": "1,088",
      "reason_for_recall": "Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.",
      "recall_initiation_date": "20230412",
      "center_classification_date": "20230503",
      "report_date": "20230510",
      "code_info": "Software version 2.03.06 and earlier and software version 2.04 and later.",
      "more_code_info": ""
    }
  ]
}