{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90559",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE,  BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ,  IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT,  QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.",
      "recall_number": "Z-1498-2022",
      "product_description": "Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative  determination of iron in human serum and (lithium heparin) plasma samples.",
      "product_quantity": "1614 units US ; 7,979 units OUS",
      "reason_for_recall": "Falsely Elevated Atellica CH Microalbumin_2 (¿ALB_2) Results due to Reagent Carryover from the Iron_2 Assay.  This defect occurs only when a microalbumin test is run immediately after an iron test.",
      "recall_initiation_date": "20220628",
      "center_classification_date": "20220804",
      "report_date": "20220810",
      "code_info": "Unique Device Identification (UDI): 00630414596402 All lots"
    }
  ]
}