      "event_id": "84756",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution. Foreign distribution worldwide.",
      "recall_number": "Z-1493-2020",
      "product_description": "Allura CV20, system code 722031",
      "product_quantity": "128",
      "reason_for_recall": "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.",
      "recall_initiation_date": "20191230",
      "center_classification_date": "20200313",
      "termination_date": "20221117",
      "report_date": "20200325",
      "code_info": "Not distributed in US.  Serial numbers  155  42  150  160  87  154  29  73  88  48  109  78  53  63  41  97  138  86  39  36  152  135  40  49  80  153  91  23  35  64  71  55  33  32  30  75  76  70  133  124  182  173  101  128  21  115  81  82  19  84  92  166  72  69  22  26  61  130  52  98  116  117  110  113  112  111  89  90  177  164  165  161  162  163  28  100  174  56  167  119  168  185  50  120  151  125  131  159  126  79  66  137  65  123  169  122  121  148  147  146  149  140  38  129  184  127  181  178  62  170  132  157  158  134  46  136  96  180  45  47  99  54  31  175  83  77  94  85",
      "more_code_info": ""
    }
  ]
}