  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98394",
      "recalling_firm": "Diagnostica Stago, Inc.",
      "address_1": "5 Century Dr",
      "address_2": "N/A",
      "postal_code": "07054-4607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BIELORUSSIE, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, GUATEMALA, INDIA, IRAQ, IRELAND, ITALY, JORDAN, KENYA, KOREA REPUBLIC OF, LITUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PALESTINIAN TERRITORY, PARAGUAY, PORTUGAL, ROMANIA, RUSSIE, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIE, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY;",
      "recall_number": "Z-1491-2026",
      "product_description": "Brand Name: STA Liatest Free Protein S  Product Name: STA Liatest Free Protein S  Model/Catalog Number (REF): 00516  Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method.  Component: Not Applicable",
      "product_quantity": "369",
      "reason_for_recall": "The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.",
      "recall_initiation_date": "20260204",
      "center_classification_date": "20260303",
      "report_date": "20260311",
      "code_info": "Model/Catalog Number (REF): 00516; UDI: (1) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (2) (01)03607450005165(11)250221(17)260531(10)272859(241)00516, (3) (01)03607450005165(11)250306(17)260630(10)273018(241)00516, (4) (01)03607450005165(11)250430(17)260831(10)273426(241)00516;  Lot numbers: (1) 271 971, (2) 272 859, (3) 273 018, (4) 273 426; Lot Code: 271 971 (01)03607450005165(11)241001(17)260131(10)271971(241)00516"
    }
  ]
}