  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plainsboro",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79636",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "311 Enterprise Dr",
      "address_2": "N/A",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-1487-2018",
      "product_description": "TCC-EZ, Total Contact Cast System,  Model No. TCC2-4051",
      "product_quantity": "250",
      "reason_for_recall": "A quarantined lot of the devices was released for sale. Devices from this lot may experience increased exothermic reaction causing higher levels of heat output which exceed the product specifications.",
      "recall_initiation_date": "20170818",
      "center_classification_date": "20180424",
      "termination_date": "20180719",
      "report_date": "20180502",
      "code_info": "Lot 1216033"
    }
  ]
}