  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Nidau",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98397",
      "recalling_firm": "Staar Surgical AG",
      "address_1": "Haupt Strasse 104",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "International distribution to the countries of India, Iran, Korea, Saudi Arabia.",
      "recall_number": "Z-1480-2026",
      "product_description": "EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5",
      "product_quantity": "7",
      "reason_for_recall": "Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.",
      "recall_initiation_date": "20260121",
      "center_classification_date": "20260302",
      "report_date": "20260311",
      "code_info": "UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988"
    }
  ]
}