  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98359",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "N/A",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.",
      "recall_number": "Z-1477-2026",
      "product_description": "Brand Name: ShockPulse-SE Lithotripsy System SPL-SR  Product Name: ShockPulse-SE Lithotripsy System - Reusable Probes  Model/Catalog Number: SPL-SR  Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.",
      "product_quantity": "602 units",
      "reason_for_recall": "Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.",
      "recall_initiation_date": "20251114",
      "center_classification_date": "20260227",
      "report_date": "20260311",
      "code_info": "Model/Catalog Number: SPL-SR; UDI: 00821925043824; Serial Numbers:  All;"
    }
  ]
}