      "status": "Terminated",
      "city": "Batesville",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64550",
      "recalling_firm": "Hill-Rom, Inc.",
      "address_1": "1069 State Route 46 East",
      "address_2": "N/A",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, IL, IN, KS, LA, MA, MI, MN, MO, NC, NH, NJ, NV, OH, OK, PA, SC, TN, UT, WA, and WI, and the countries of Canada, Australia, Austria, Belgium, Switzerland, Germany Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Netherlands, Norway, New Zealand, Portugal, Sweden, and Turkey.",
      "recall_number": "Z-1474-2013",
      "product_description": "SlingBar 360.    Designed to meet the needs for lifting humans.",
      "product_quantity": "16",
      "reason_for_recall": "Retrospective review found 4 sling bars that may present a potential for impalement if a patient, caregiver or bystander were to fall onto the hooks based on a simulated test method. In effort to eliminate any risk of injury, Hill-Rom  is requesting products be removed from field use and  replaced with an updated design.",
      "recall_initiation_date": "20130430",
      "center_classification_date": "20130605",
      "termination_date": "20141128",
      "report_date": "20130612",
      "code_info": "Product # 3156013"
    }
  ]
}