  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lexington",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92012",
      "recalling_firm": "FUJIFILM Healthcare Americas Corporation",
      "address_1": "81 Hartwell Ave Ste 300",
      "address_2": "N/A",
      "postal_code": "02421-3160",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1470-2023",
      "product_description": "FujiFilm FDR AQRO (Model DR-XD1000)- A digital mobile X-ray system intended for use in general purpose radiography for generating radiographic images of human anatomy, including adult, pediatric, and neonatal exams",
      "product_quantity": "77 units",
      "reason_for_recall": "Arm which holds the tube head in place is susceptible to breaking, if the device breaks it could potentially cause serious harm or injury to the patient and/or user.",
      "recall_initiation_date": "20230227",
      "center_classification_date": "20230427",
      "report_date": "20230503",
      "code_info": "UDI: (01)04547410425505(11)200701. Serial numbers with the last 4 digits ending between ****0001 to ****1993."
    }
  ]
}