  "results": [
    {
      "status": "Ongoing",
      "city": "Hazelwood",
      "state": "MO",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94264",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "address_2": "N/A",
      "postal_code": "63042-2320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-1464-2024",
      "product_description": "VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility):    a) AST-N334, REF 418984;   b) AST-GN95, REF 421982;   c) AST-GN99, REF 423102;   d) AST-N390, REF 423340;   e) AST-N391, REF 423341;   f) AST-N392, REF 423342;   g) AST-N395, REF 423491;   h) AST-N401, REF 423643;   i) AST-N402, REF 423644;   j) AST-N404, REF 423664;   k) AST-N802, REF 423706;   l) AST-N405, REF 423864;   m) AST-N417, REF 423880;   n) AST-N408, REF 423924;   o) AST-N409, REF 423925;   p) AST-N420, REF 424039;   q) AST-N423, REF 424042;   r) AST-N422, REF 424056;   s) AST-N436, REF 424440;   t) AST-N809, REF 424703;   u) AST-N806, REF 424709;   v) AST-N807, REF 424710;   w) AST-N808, REF 424711;   x) AST-N810, REF 424712;   y) AST-N812, REF 424721",
      "product_quantity": "49215 units",
      "reason_for_recall": "Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results.  The risk is present for isolates that have a result of MIC=0.5, 1, 2.",
      "recall_initiation_date": "20240314",
      "center_classification_date": "20240425",
      "report_date": "20240501",
      "code_info": "a) AST-N334, REF 418984, UDI/DI 3573026490720, Lot/Serial Numbers:  7742790503, 7742805403, 7742828503, 7742830203;   b) AST-GN95, REF 421982, UDI/DI 3573026565923, Lot/Serial Numbers:   6852740103, 6852751503, 6852778503, 6852811503;   c) AST-GN99, REF 423102, UDI/DI 3573026608132, Lot/Serial Numbers:  6892721103, 6892727403, 6892740503, 6892740403, 6892743503, 6892761503, 6892770403, 6892790103, 6892799403, 6892799103, 6892810103, 6892818103, 6892831503;   d) AST-N390, REF 423340, UDI/DI 3573026612177, Lot/Serial Numbers:  1402832104, 1402761504;   e) AST-N391, REF 423341, UDI/DI 3573026612269, Lot/Serial Numbers:  1412831104, 1412754504;   f) AST-N392, REF 423342, UDI/DI 3573026612283, Lot/Serial Numbers:  1422819204;   g) AST-N395, REF 423491, UDI/DI 3573026615802, Lot/Serial Numbers:  1452725204;   h) AST-N401, REF 423643, UDI/DI 3573026616816, Lot/Serial Numbers:  1512776404, 1512789204, 1512811104, 1512831104, 1512720404, 1512729404, 1512729104, 1512748504, 1512754104, 1512785404;   i) AST-N402, REF 423644, UDI/DI 3573026616830, Lot/Serial Numbers:  1522720504, 1522733404;   j) AST-N404, REF 423664, UDI/DI 3573026617356, Lot/Serial Numbers:  1542819204;   k) AST-N802, REF 423706, UDI/DI 3573026617752, Lot/Serial Numbers: 0422750104, 0422763104, 0422764504, 0422782404, 0422820504, 0422820404;   l) AST-N405, REF 423864, UDI/DI 3573026621360, Lot/Serial Numbers:  1552757504, 1552792404, 1552800404, 1552820104, 1552733104, 1552741504, 1552757404;   m) AST-N417, REF 423880, UDI/DI 3573026621704, Lot/Serial Numbers: 0192763504, 0192819404;   n) AST-N408, REF 423924, UDI/DI 3573026622725, Lot/Serial Numbers:   1582810504, 1582820104, 1582734404, 1582742504, 1582764404, 1582782504;   o) AST-N409, REF 423925, UDI/DI 3573026622749, Lot/Serial Numbers:  1592811104, 1592746404;    p) AST-N420, REF 424039, UDI/DI 3573026625283, Lot/Serial Numbers:  0462810404;    q) AST-N423, REF 424042, UDI/DI 3573026625337, Lot/Serial Numbers:  0502750404, 0502825504;    r) AST-N422, REF 424056, UDI/DI 3573026625610, Lot/Serial Numbers:  0482819504;    s) AST-N436, REF 424440, UDI/DI 3573026631666, Lot/Serial Numbers:  0672734104, 0672754104;   t) AST-N809, REF 424703, UDI/DI 3573026635978, Lot/Serial Numbers:  0962799404;    u) AST-N806, REF 424709, UDI/DI 3573026636067, Lot/Serial Numbers:  0932729404;   v) AST-N807, REF 424710, UDI/DI 3573026636081, Lot/Serial Numbers:  0942723404;   w) AST-N808, REF 424711, UDI/DI 3573026636104, Lot/Serial Numbers: 0952758104;    x) AST-N810, REF 424712, UDI/DI 3573026636128, Lot/Serial Numbers:  0972729504;    y) AST-N812, REF 424721, UDI/DI 3573026636296, Lot/Serial Numbers:  0992777104, 0992776504, 0992821404",
      "more_code_info": ""
    }
  ]
}