  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98333",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1463-2026",
      "product_description": "Medline Kits containing Tego Connectors:  Medline IR Pack, SKU DYNJ67205C  Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200",
      "product_quantity": "516",
      "reason_for_recall": "Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.",
      "recall_initiation_date": "20260108",
      "center_classification_date": "20260224",
      "report_date": "20260304",
      "code_info": "IR PACK UDI-DI (ea) 10195327621780 UDI-DI (case) 40195327621781 Kit lots 24BBP627 24DBO029 24FBI148 24JBA858 24KBD925 24LBG978 25ABQ002 25EBV162;  NO CATHETER DIALYSIS INSERTION UDI-DI (ea) 10195327675219 UDI-DI (case) 40195327675210 Kit Lots 24EBM915 24GBC287"
    }
  ]
}