  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "7301 Georgetown Rd.",
      "reason_for_recall": "Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the",
      "address_2": "Suite 150",
      "product_quantity": "33,543 all kits",
      "code_info": "Pump Part 510613 , Lot 11-100815 - Lot 12-10101060 found in: GoBlock Kit 510612, Lot 11-100817 to 13-101475",
      "center_classification_date": "20130606",
      "distribution_pattern": "US Nationwide distribution.",
      "state": "IN",
      "product_description": "Part 510613 Disposable  Infusion Pump, ...Elastomeric  Pump 300mL (150 mL side)( 6mL/hr ),...  found in the following kits  GoBlock Kit 510612,    Disposable Infusion Pump for drug delivery. Symbios GOBlock System is a disposable, self contained infusion system.",
      "report_date": "20130612",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Symbios Medical Products, LLC",
      "recall_number": "Z-1461-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65192",
      "termination_date": "20140307",
      "recall_initiation_date": "20130510",
      "postal_code": "46268",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}