      "status": "Terminated",
      "city": "Lake Oswego",
      "state": "OR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87549",
      "recalling_firm": "BIOTRONIK Inc",
      "address_1": "6024 Jean Rd",
      "address_2": "N/A",
      "postal_code": "97035-5571",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide  OUS: Worldwide",
      "recall_number": "Z-1453-2021",
      "product_description": "BIOTRONIK IPERIA 7, DR-T DF4 ProMRI, REF 392423, UDI: 04035479128913 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.",
      "product_quantity": "2881 UNITS",
      "reason_for_recall": "There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)",
      "recall_initiation_date": "20210308",
      "center_classification_date": "20210421",
      "termination_date": "20230425",
      "report_date": "20210428",
      "code_info": "Serial Number  60826231  60826274  60826822  60834578  60834590  60834592  60834595  60834602  60834610  60834618  60834627  60834628  60834640  60835705  60835712  60836005  60836015  60836158  60836479  60836767  60836769  60836770  60836771  60836775  60836776  60836777  60836779  60836782  60836787  60836788  60836802  60836809  60836812  60836824  60836830  60836935  60836941  60836944  60836963  60836979  60836985  60837879  60837889  60837892  60837919  60837922  60837924  60837928  60837932  60838019  60838020  60838056  60838060  60838064  60838082  60838095  60838106  60838107  60838110  60838382  60838656  60838668  60838670  60838673  60838674  60838680  60838683  60838689  60838704  60838941  60839108  60839144  60839263  60839808  60839811  60839827  60839830  60839833  60839838  60839841  60839845  60839857  60839863  60839868  60839871  60839873  60839874  60839875  60840403  60840404  60840406  60840411  60840414  60840420  60840421  60840422  60840444  60840454  60840459  60840460  60840638  60841087  60841090  60841146  60841153  60841154  60841156  60841157  60841159  60841160  60841165  60841166  60841167  60841170  60841179  60841186  60841193  60841195  60841196  60841202  60841206  60841208  60841210  60841215  60841220  60841226  60841230  60841235  60841238  60841241  60841365  60841370  60841505  60841509  60841512  60841968  60841974  60842003  60842004  60843951  60843978  60843984  60843986  60843987  60843988  60843999  60844007  60844016  60844018  60844342  60844344  60844345  60844346  60844349  60844355  60844357  60844360  60844361  60844368  60844370  60844374  60844375  60844382  60844386  60844388  60844389  60844397  60844400  60844421  60844430  60844699  60844700  60844701  60844705  60844727  60844728  60844730  60844732  60844738  60844740  60844741  60844742  60844743  60844744  60844745  60845578  60845585  60845586  60845593  60845597  60845599  60845601  60845602  60845603  60845605  60845606  60846068  60846070  60846072  60846075  60846078  60846083  60846085  60846086  60846088  60846089  60846090  60846100  60846105  60846130  60846310  60846871  60846872  60846873  60846880  60846881  60846955  60846956  60846972  60846980  60847223  60848255  60848256  60848258  60848259  60848260  60848262  60848266  60848268  60848269  60848284  60848290  60848292  60848293  60848298  60848302  60848305  60848309  60848310  60848325  60848326  60848328  60848334  60848339  60848453  60848454  60848455  60848456  60848466  60848470  60848476  60848477  60848481  60848484  60848486  60848488  60848490  60848493  60848495  60848503  60849278  60849280  60849287  60849288  60849290  60849291  60849292  60849294  60849295  60849296  60849297  60849299  60849302  60849309  60849310  60849313  60849314  60849317  60849320  60849321  60849323  60849325  60849326  60849327  60849333  60849417  60849419  60849423  60849424  60849425  60849430  60849437  60849438  60849444  60849445  60849446  60849447  60849448  60849449  60849451  60849452  60850393  60850408  60850409  60850410  60850416  60850419  60850424  60850432  60850438  60850442  60850446  60850469  60850475  60850496  60850524  60851582  60851583  60851586  60851588  60851591  60851592  60851593  60851594  60851595  60851599  60851606  60851614  60851616  60851617  60851639  60851640  60851642  60851646  60851648  60851652  60851656  60851657  60851660  60851661  60851663  60851667  60851686  60851687  60851688  60851689  60851691  60851692  60851699  60851704  60851706  60851712  60851714  60851716  60851717  60851718  60851720  60851721  60851726  60851728  60851730  60851732  60851733  60851919  60851920  60851921  60851926  60851928  60851930  60851931  60851932  60851934  60851935  60851937  60851938  60851941  60851945  60851948  60851949  60851958  60851959  60851969  60851970  60851991  60851994  60851998  60852269  60852273  60852275  60852277  60852280  60852282  60852285  60852287  60852288  60852301  60852306  60852307  60852309  60852310  60852312  60852314  60852320  60852324  60852329  60852330  60852331  60852335  60852336  60852342  60852348  60852912  60852913  60852915  60852918  60852919  60852920  60852922  60852923  60852924  60852926  60852927  60852928  60852930  60852934  60852946  60853099  60853103  60853104  60853105  60853108  60853109  60853114  60853116  60853117  60853119  60853122  60853124  60853127  60853128  60853132  60853133  60853134  60853351  60853358  60853361  60853364  60854280  60854283  60854284  60854285  60854287  60854292  60854296  60854521  60854525  60854527  60854541  60854543  60854545  60854547  60854551  60854552  60854553  60854554  60854557  60854558  60854559  60854560  60854563  60854565  60854566  60854568  60854570  60854572  60854573  60854574  60854575  60854576  60854577  60854578  60854579  60854582  60854584  60854585  60854586  60854682  60854688  60854692  60854693  60854696  60854697  60854700  60854701  60854710  60854711  60854712  60854713  60854721  60854723  60854724  60854725  60854729  60854730  60855423  60855425  60855427  60855428  60855429  60855433  60855434  60855437  60855443  60855446  60855448  60855449  60855451  60855452  60855453  60855459  60855466  60855468  60855469  60855470  60855474  60855475  60855477  60855479  60855480  60855481  60855482  60855849  60855851  60855852  60855853  60855855  60855856  60855857  60855858  60855920  60855922  60855924  60855925  60855927  60855931  60855934  60855935  60855936  60855937  60855938  60855940  60855943  60855945  60855947  60855948  60855953  60855956  60855957  60856190  60856193  60856195  60856201  60856205  60856209  60856210  60856212  60856215  60856216  60856217  60856218  60856749  60856750  60856752  60856756  60856758  60856759  60856761  60856762  60856767  60856771  60856773  60856781  60856783  60856787  60856790  60856791  60856796  60856802  60856808  60856809  60856970  60856971  60856976  60856977  60857166  60857167  60857175  60857177  60857178  60857179  60857180  60857183  60857184  60857186  60857189  60857193  60857195  60857196  60857199  60857200  60857202  60857203  60857205  60857212  60857217  60857470  60857472  60857473  60857476  60857477  60857478  60857487  60857489  60857493  60857499  60857501  60857505  60857506  60857508  60857510  60857511  60857513  60857515  60857517  60857520  60857522  60857526  60857527  60857528  60857529  60857640  60857641  60857642  60857644  60857645  60858171  60858172  60858173  60858174  60858175  60858180  60858183  60858185  60858186  60858187  60858189  60858193  60858194  60858196  60858197  60858198  60858199  60858200  60858202  60858203  60858204  60858205  60858206  60858207  60858211  60858217  60858219  60858220  60858221  60858224  60858225  60858228  60858229  60858231  60858232  60858236  60858238  60858375  60858376  60858378  60858382  60858383  60858384  60858389  60858391  60858392  60858393  60858395  60858396  60858401  60858403  60858405  60858406  60858408  60858410  60858411  60858412  60858413  60858414  60858415  60858416  60858417  60858418  60858419  60858420  60858494  60858495  60858497  60858498  60858499  60858500  60858501  60858502  60858503  60858505  60858506  60858508  60858509  60858510  60858515  60858516  60858517  60858519  60858522  60858523  60858526  60858527  60858528  60858533  60858535  60858536  60858538  60858540  60858542  60858544  60858545  60858546  60858732  60858734  60858735  60858736  60858737  60858741  60858743  60858746  60858747  60858750  60859423  60859426  60859427  60859428  60859430  60859432  60859433  60859437  60859438  60859439  60859440  60859441  60859720  60859722  60859724  60859727  60859728  60859730  60859731  60859732  60859734  60859736  60859739  60859740  60859742  60859744  60859745  60859812  60859813  60859815  60859816  60859817  60859819  60859822  60859823  60859829  60859831  60859832  60859836  60859837  60859838  60859840  60859872  60859873  60859875  60859876  60859877  60859878  60859881  60859882  60859883  60859885  60859886  60859887  60859889  60859890  60859962  60859965  60859966  60859969  60860086  60860088  60860090  60860144  60860146  60860148  60860149  60860153  60860154  60860155  60860157  60860158  60860160  60860161  60860163  60860164  60860165  60860166  60860170  60860177  60860178  60860182  60860195  60860197  60860745  60860746  60860748  60860753  60860754  60860757  60860759  60860760  60860762  60860763  60860765  60860769  60860774  60860775  60860779  60860781  60860783  60860785  60860787  60860789  60860790  60860792  60860794  60863140  60863142  60863143  60863144  60863146  60863148  60863149  60863150  60863152  60863153  60863154  60863155  60863156  60863157  60863159  60863162  60863163  60863164  60863165  60863167  60863168  60863169  60863170  60863171  60863172  60863173  60863174  60863175  60863178  60863179  60863180  60863181  60863182  60863183  60863189  60863547  60863548  60863549  60863555  60863561  60863562  60863563  60863564  60863566  60863567  60863568  60863570  60863571  60863574  60863575  60863576  60863577  60863578  60863579  60863581  60863582  60863583  60863584  60863745  60863746  60863748  60863750  60863751  60863753  60863755  60863757  60863759  60863760  60863761  60863762  60863764  60863765  60863766  60863767  60863769  60863770  60863772  60863773  60863774  60864262  60864264  60864266  60864267  60864268  60864270  60864271  60864278  60864281  60864282  60864283  60864284  60864285  60864286  60864287  60864288  60864291  60864295  60864296  60864298  60864300  60864301  60864302  60864306  60864307  60864981  60864982  60864983  60864984  60864985  60864986  60864988  60864989  60864990  60864993  60864995  60864996  60865000  60865001  60865003  60865006  60865008  60865009  60865110  60865111  60865113  60865117  60865118  60865120  60865123  60865125  60865126  60865127  60865131  60865132  60865136  60865137  60865139  60865290  60865291  60865292  60865295  60865296  60865297  60865298  60865299  60865300  60865302  60865303  60865304  60865305  60865306  60865307  60865308  60865309  60865742  60865743  60865744  60865746  60865747  60865748  60865750  60865756  60865762  60865764  60865765  60865767  60865770  60865771  60865772  60865773  60865775  60865776  60865778  60865779  60865780  60865781  60865782  60865785  60865786  60865787  60865788  60865790  60865791  60865792  60865793  60865794  60865795  60865796  60865799  60865800  60865802  60865803  60865804  60865806  60865809  60865810  60865811  60865843  60865844  60865846  60865847  60865849  60865852  60865853  60865854  60865855  60865856  60865857  60865858  60865859  60865861  60865862  60865863  60865868  60865871  60865872  60865875  60865877  60865879  60865882  60865885  60865886  60865887  60865889  60865891  60866182  60866186  60866194  60866196  60866197  60866198  60866200  60866201  60866203  60866210  60866211  60866933  60866934  60866936  60866938  60866947  60866948  60866950  60866951  60866953  60866954  60866956  60866958  60866965  60866966  60866967  60866968  60867046  60867047  60867049  60867051  60867055  60867058  60867059  60867061  60867063  60867066  60867069  60867072  60867074  60867075  60867229  60867230  60867232  60867234  60867236  60867238  60867240  60867241  60867243  60867244  60867247  60867248  60867251  60867252  60867253  60867261  60867263  60867264  60867265  60867268  60867272  60867274  60867431  60867434  60867435  60867436  60867437  60867440  60867446  60867458  60867459  60868335  60868341  60868346  60868348  60868356  60868357  60868362  60868363  60868364  60868365  60868366  60868367  60868370  60868380  60868387  60868389  60868394  60868396  60868402  60868407  60868411  60868413  60868422  60868435  60868438  60868439  60868450  60868451  60868453  60868455  60868469  60868470  60868473  60869646  60869647  60869649  60869651  60869660  60869661  60869662  60869665  60869668  60869669  60869672  60869673  60869675  60869676  60869677  60869678  60869681  60869682  60869683  60869684  60869687  60869693  60869695  60869846  60869847  60869848  60869849  60869851  60869855  60869858  60869861  60869864  60869867  60869868  60869872  60869874  60869879  60869888  60869889  60869890  60869891  60869893  60870386  60870388  60870407  60870425  60870428  60870430  60870437  60870572  60870573  60870575  60870577  60870578  60870583  60870585  60870598  60870604  60870606  60870607  60870612  60870614  60870616  60870618  60870621  60870625  60870626  60870628  60870630  60870631  60870632  60871444  60871447  60871451  60871454  60871464  60871473  60871477  60871479  60871481  60871483  60871484  60871488  60871501  60871502  60871504  60871505  60871508  60871509  60871510  60871516  60871517  60871520  60871525  60871526  60871529  60872268  60872270  60872271  60872274  60872275  60872276  60872285  60872289  60872292  60872296  60872298  60872305  60872306  60872308  60872309  60872310  60872313  60872406  60872409  60872412  60872416  60872418  60872419  60872420  60872424  60872435  60872446  60872448  60872449  60872587  60872588  60872595  60872601  60872604  60872605  60872877  60872878  60872881  60872885  60872886  60872888  60872889  60872890  60872893  60872897  60872899  60872900  60872901  60872903  60872904  60872912  60873517  60873519  60873531  60873533  60873544  60873549  60873550  60873552  60873554  60873557  60873558  60873560  60873563  60873569  60873574  60873576  60873584  60873586  60873690  60873692  60874845  60874846  60874854  60874884  60874897  60875702  60875703  60875707  60875710  60875712  60875714  60875717  60875718  60875722  60875728  60875730  60875732  60875734  60875735  60875736  60875739  60875740  60875741  60875935  60875948  60875950  60875951  60875952  60875953  60875954  60875956  60875958  60875966  60875970  60875971  60875977  60875981  60875988  60875991  60875993  60876154  60876158  60876162  60876163  60876165  60876168  60876170  60876172  60876173  60876174  60876176  60876177  60876181  60876182  60877734  60877735  60877740  60877742  60877743  60877744  60877749  60877750  60877751  60877778  60877779  60877792  60877793  60877798  60877799  60877803  60877804  60877805  60877806  60878655  60878659  60878668  60878676  60878679  60878680  60878683  60878689  60878715  60878951  60878956  60879758  60879763  60879766  60879767  60879768  60879772  60879851  60879852  60879857  60879860  60879861  60879862  60879865  60879867  60879872  60879875  60879878  60879880  60879882  60879895  60879896  60879903  60879904  60879909  60879910  60879915  60879918  60879938  60880349  60880354  60880366  60880373  60880381  60880382  60880385  60880391  60880392  60880393  60880394  60880399  60880406  60880439  60880450  60880455  60880457  60880809  60880819  60880830  60880831  60880834  60880837  60880839  60880842  60880844  60880845  60880847  60880853  60880855  60880858  60880861  60880864  60880865  60880866  60880867  60880868  60880869  60880871  60881364  60881371  60881375  60881379  60881380  60881381  60881599  60881607  60881610  60881611  60881613  60881614  60881615  60881657  60881658  60881660  60881661  60881663  60881670  60881675  60881678  60881683  60881687  60881688  60881692  60881698  60881702  60881704  60881707  60881708  60881711  60881712  60881716  60881719  60881721  60881722  60882007  60882008  60882009  60882013  60882014  60882015  60882017  60882018  60882019  60882020  60882021  60882023  60882031  60882033  60882039  60882042  60882043  60882044  60882046  60882049  60882053  60883328  60883329  60883333  60883335  60883340  60883343  60883345  60883350  60883351  60883356  60883358  60883360  60883363  60883364  60883365  60883366  60883372  60883374  60883377  60883379  60883380  60883381  60883385  60883386  60883391  60883397  60883398  60883399  60883402  60883405  60883406  60883408  60883410  60883411  60883413  60883415  60883416  60883417  60883419  60883420  60883421  60883422  60883423  60883425  60883688  60883689  60883690  60883691  60883692  60883694  60883696  60884180  60884181  60884198  60884204  60884206  60884207  60884214  60884432  60884434  60884435  60884436  60884437  60884438  60884442  60884443  60884444  60884445  60884446  60884448  60884449  60884450  60884451  60884452  60884453  60884455  60884460  60884461  60884463  60884464  60884465  60884467  60884469  60884470  60884471  60884475  60884477  60884478  60884479  60884480  60884482  60884483  60884484  60884485  60884487  60884488  60884490  60884491  60884492  60884493  60884494  60884495  60884497  60884498  60884499  60884500  60884501  60884502  60884503  60884504  60884505  60884506  60884507  60884508  60884509  60884510  60884511  60884512  60884514  60884515  60884516  60884518  60884519  60884520  60884522  60884527  60884530  60885401  60885403  60885405  60885410  60885411  60885412  60885413  60885414  60885415  60885416  60885417  60885418  60885420  60885421  60885427  60885429  60885435  60885436  60885438  60885439  60885440  60885441  60885442  60885445  60885446  60885447  60885448  60885449  60885451  60885453  60885454  60885455  60885456  60885459  60885460  60885461  60885462  60885463  60885465  60885470  60885622  60885623  60885624  60885626  60885628  60885631  60885633  60885634  60885636  60885641  60885642  60885643  60885646  60885647  60885648  60885649  60885650  60885652  60885655  60885660  60885662  60885665  60885667  60885669  60885670  60885673  60885674  60885675  60885676  60885677  60885678  60885679  60885680  60908006  60908008  60908011  60908013  60908014  60908017  60908023  60908025  60908029  60908033  60908040  60908043  60908045  60908047  60908048  60908055  60908746  60908750  60908751  60908755  60908757  60908758  60908761  60908762  60908763  60908764  60908766  60908767  60908768  60908769  60908771  60908772  60908774  60909325  60909326  60909328  60909331  60909333  60909334  60909337  60909339  60909341  60909342  60909344  60909351  60910242  60910247  60910250  60910254  60910255  60910256  60910258  60910261  60910264  60910272  60910276  60910279  60910284  60910286  60910294  60910297  60910298  60910299  60910301  60910306  60910309  60910311  60912746  60912760  60912761  60912764  60913587  60913588  60913590  60913591  60913592  60913594  60913595  60913596  60913598  60913599  60913600  60913603  60913604  60913605  60913607  60913612  60913613  60913614  60913616  60913825  60913828  60913829  60913831  60913833  60913834  60913836  60913839  60913843  60913844  60913849  60913850  60913852  60913854  60913858  60913861  60913870  60914049  60914052  60914057  60914059  60914060  60914062  60914064  60914065  60914066  60914068  60914073  60914075  60914076  60914078  60914082  60914083  60914084  60914085  60914087  60914088  60914089  60914095  60914638  60914639  60914643  60914644  60915340  60915342  60915344  60915345  60915352  60915357  60915358  60915360  60915361  60915363  60915365  60915366  60915369  60915629  60915631  60915636  60915637  60915638  60915639  60915640  60915641  60915642  60915644  60915646  60915648  60915654  60915656  60915658  60915662  60915664  60915665  60915667  60915669  60915671  60915673  60915674  60915675  60915678  60915680  60915682  60915683  60915685  60915686  60915691  60915692  60915693  60915695  60915696  60916468  60916472  60916473  60916474  60916475  60916478  60916480  60916482  60916483  60916486  60916488  60916491  60916493  60916494  60916495  60916497  60916498  60916499  60916501  60916502  60916503  60916633  60916636  60916641  60916642  60916643  60916645  60916646  60916647  60916648  60916649  60916650  60916653  60916655  60916657  60916658  60916659  60916662  60916664  60916666  60916667  60916668  60916669  60916670  60916671  60916672  60916673  60916675  60916677  60916678  60916679  60916680  60916683  60916684  60916686  60916687  60916688  60916690  60916693  60916695  60916697  60916698  60916700  60916702  60916703  60916705  60916706  60916708  60918201  60918203  60918205  60918208  60918210  60918401  60918403  60918404  60918406  60918407  60918409  60918410  60918411  60918412  60918413  60918414  60918415  60918416  60918418  60918420  60918421  60918422  60918423  60918424  60918425  60918426  60918427  60918429  60918430  60918432  60918433  60918436  60918437  60918439  60918440  60918442  60918443  60918445  60918446  60918447  60918449  60918451  60918453  60918455  60918457  60918458  60918459  60918461  60918462  60918463  60918465  60918466  60918467  60918468  60918469  60918470  60918471  60918472  60918473  60918474  60918475  60918477  60918478  60918481  60918482  60918483  60918484  60918485  60918486  60918488  60918489  60918490  60918778  60918779  60918780  60918781  60918782  60918783  60918785  60918787  60918789  60918791  60918902  60918903  60919779  60919780  60919781  60919782  60919783  60919784  60919788  60919790  60919791  60919797  60919801  60919802  60919806  60919807  60919811  60919814  60919818  60920199  60920201  60920202  60920206  60920207  60920208  60920209  60920211  60920213  60920214  60920216  60920217  60920220  60920221  60920223  60920226  60920228  60920230  60920233  60920237  60920238  60920241  60920242  60920243  60920244  60920245  60920246  60920707  60920708  60920709  60920710  60920712  60920714  60920715  60920717  60920718  60920728  60920729  60920732  60920733  60920736  60920739  60920741  60920742  60920744  60920752  60920755  60922117  60922120  60922122  60922125  60922126  60922127  60922128  60922129  60922130  60922131  60922132  60922135  60922141  60922142  60922145  60922151  60922152  60922161  60922163  60922164  60922169  60922170  60922173  60922177  60922178  60922182  60922183  60922190  60924897  60924899  60924900  60924901  60924904  60924907  60924908  60924909  60924912  60924913  60924914  60924919  60924921  60924924  60924935  60924936  60924937  60924939  60924941  60924943  60924945  60924948  60925961  60925963  60925966  60925968  60925978  60925980  60925984  60925986  60927586  60927594  60929086  60929094  60929100  60929105  60929108  60929897  60929899  60929905  60929907  60929913  60929915  60929930  60929933  60929935  60929937  60929939  60929943  60929945  60929956  60930138  60930139  60930141  60930142  60930144  60930146  60930148  60930155  60930158  60931041  60931042  60931044  60931045  60931046  60931048  60931050  60931073  60931075  60934371  60934373  60934377  60934379  60934384  60934385  60934386  60934388  60934391  60934392  60934396  60934402  60934408  60936166  60937913  60937918  60937919  60937921  60937928  60937934  60937937  60937942  60937945  60937955  60937956  60937958  60938830  60938841  60938842  60938850  60938852  60938854  60938856  60938858  60940028  60940031  60940034  60940036  60940040  60940045  60940046  60940051  60940067  60941221  60941223  60941230  60941236  60941237  60941238  60941239  60941242  60941246  60941247  60943186  60943187  60943192  60943194  60943195  60943196  60943198  60943199  60943200  60943201  60943203  60943205  60943206  60943210  60944415  60947747  60947749  60947751  60947752  60947754  60948183  60948186  60948187  60948298  60948299  60948306  60948310  60948311  60948312  60948313  60948316  60951083  60951084  60951087  60951088  60951090  60951093  60951094  60951095  60951099  60952174  60952179  60952183  60952184  60952187  60952188  60952189  60952772  60952773  60952778  60952780  60952781  60952782  60952783  60952785  60952787  60952789  60952792  60952795  60952796  60952859  60952860  60952863  60952864  60952873  60952874  60952886  60952888  60952889  60952890  60952893  60952894  60952895  60952896  60952897  60952901  60954393  60954403  60954404  60954412  60954420  60954421  60954424  60954425  60954426  60954431  60954432  60956565  60956568  60956571  60956573  60956578  60956579  60956581  60956583  60956585  60956586  60956587  60956591  60956592  60956593  60956598  60956601  60956603  60956608  60956611  60956616  60956617  60956620  60956622  60956623  60958214  60958215  60958221  60958222  60958223  60958224  60958227  60958229  60958235  60958240  60958245  60958258  60958522  60958526  60958530  60958534  60958542  60958545  60958547  60958549  60958550  60958551  60958557  60958559  60958561  60958563  60958565  60958567  60958568  60958570  60959725  60959748  60959752  60959757  60959758  60959762  60959765  60959999  60960000  60960006  60960013  60960017  60960020  60960023  60960024  60960027  60960983  60960985  60960990  60961292  60961299  60961300  60961301  60961305  60961306  60961307  60963181  60963186  60963188  60963189  60963190  60963191  60963192  60963195  60963196  60963203  60963216  60963221  60963224  60964418  60964420  60964425  60964427  60964430  60964431  60964432  60964433  60964435  60964436  60964440  60964442  60965109  60965838  60965841  60965843  60965844  60965848  60965858  60965860  60965862  60966103  60966118  60966122  60967401  60967404  60967405  60968302  60968304  60968306  60968311  60968316  60968318  60969485  60969487  60969491  60969500  60975354  60975359  60975366  60976183  60976186  60977863  60977866  60977871  60980589  60980605  60980607  60980615  60980619  60980620  60980621  60980623  60980624  60980627  60980628  60980629  60980655  60980656  60980657  60980658  60980662  60980667  60980677  60980682  60980684  60981033  60982949  60982951  60982954  60982956  60982957  60982977  60982978  60982980  60982981  60982985  60982986  60983318  60983326  60984294  60984295  60984298  60984300  60984301  60984306  60984307  60985809  60985811  60985813  60985814  60985815  60985816  60985817  60985829  60985837  60985838  60986132  60986142  60986147  60986155  60986156  60986157  60986158  60986159  60986160  60986162  60986163  60987343  60987347  60987349  60987351  60987357  60987360  60987372  60987375  60987377  60990615  60990619  60990654  60990677  60990772  60990783  60990789  60996068  60996070  60996071  60996072  60996073  60996074  60996075  60996076  60996079  60996080  60996081  60996082  60996083  60996086  60996087  60996088  60996090  60996093  60996094  60996096  60996097  60996098  60996099  60996100  60996101  60996103  60996104  60996105  60996106  60996794  60996796  60996798  60996799  60996801  60996802  60996803  60996807  60996808  60996809  60996811  60996812  60996816  60996817  60998270  60998272  60998274  60998275  60998277  60998278  60998280  60998281  60998282  60998283  60998284  60998286  60998287  60998288  60998289  60998290  60998291  60998292  60998293  60998294  60998295  60998296  60998297  60998298  60998299  60998300  60998301  60998302  60998303  60998305  60998306  60998307  60998308  60998927  60998929  60998931  60998932  60998933  60998935  60998936  60999360  60999362  60999364  60999365  60999367  60999369  60999371  60999374  60999383  60999385  60999386  60999389  60999540  60999542  60999543  60999544  60999545  60999547  60999551  60999552  60999555  60999556  60999557  60999569  61001358  61001360  61001361  61001362  61001363  61003010  61003011  61003012  61003013  61003014  61003015  61003018  61003019  61003020  61003021  61003022  61003023  61003175  61003176  61003177  61003178  61003180  61003182  61009378  61009633  61009636  61009638  61009639  61009641  61009643  61009644  61009645  61009647  61009648  61009649  61009651  61009652  61009653  61009654  61009656  61009657  61009658  61009660  61009661  61010452  61010453  61010454  61010455  61010459  61010460  61010461  61010464  61010465  61017378  61017380  61017382  61017383  61017384  61017385  61017386  61017387  61017389  61017390  61017391  61017393  61017395  61017396  61017397  61018538  61018541  61018542  61018545  61018547  61019269  61019272  61019274  61019275  61019279  61019280  61019282  61019287  61019811  61019815  61022412  61022416  61022417  61022418  61022419  61027821  61029797  61029805  61031266  61032329  61034659  61034660  61034669  61034673  61034674  61034677  61034679  61034680  61034682  61035522  61035525  61035527  61035528  61035529  61035530  61035531  61035533  61035534  61035537  61035538  61035542  61035543  61035545  61035546  61035549  61035550  61035551  61035554  61035555  61036621  61036622  61036624  61036625  61036628  61036632  61036633  61036636  61036637  61036638  61036648  61036649  61036650  61036651  61036653  61036655  61036657  61036660  61037862  61037871  61037872  61037873  61037874  61037882  61037898  61039694  61039695  61039698  61039699  61039700  61039705  61039706  61039707  61039708  61039712  61039716  61039717",
      "more_code_info": ""
    }
  ]
}