  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Komorniki",
      "state": "",
      "country": "Poland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84881",
      "recalling_firm": "ARJOHUNTLEIGH POLSKA Sp. z.o.o.",
      "address_1": "Ui. Ks. Piotra Wawrzyniaka 2",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: CA, CO, CT, GA, MA, MD, MI, NC, NJ, NY, OH, OK, TX, VA, WI  OUS: Australia Austria, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iceland, India, Ireland, Italy, Japan, Korea, Luxembourg, Netherlands, Netherlands/CW, New Zealand, Norway, Philippines, Poland, Romaina, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, United Arab Emirates, United Kingdom",
      "recall_number": "Z-1452-2020",
      "product_description": "Concerto & Basic  Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and     Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,",
      "product_quantity": "669 trollys",
      "reason_for_recall": "The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a  patient fall. The patient falling could lead to various types of   injuries.",
      "recall_initiation_date": "20190822",
      "center_classification_date": "20200306",
      "termination_date": "20220215",
      "report_date": "20200318",
      "code_info": "Concert & Basic Shower trolleys Serial Numbers: P0512263 to P0551621, P0279164, SEE0643339, SEE0605124,P0249502, SEE045073, SEE0520032, 1106000338,  SEE0348065, P0422951, P0232526",
      "more_code_info": ""
    }
  ]
}