  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Road",
      "reason_for_recall": "An inspection shows several IATDs (Installation Acceptance Test could not be located or different templates have been used",
      "address_2": "",
      "product_quantity": "66 units",
      "code_info": "Serial Numbers:  335292  351085  351326  348891  348895  353313  358016  359426  370503  354382  354749  369640  380748  356820  362256  357699  362257  357255  358489  380361  363543  362719  363669  363544  363305    363538  365150  364211  379698  380749  375725  375727  375729  377144  381880  381881  380798/SN07000617  381877  383767/SN07000639  392883/SN08000099  394519/SN08000135  421035/SN09000031  396577/SN08000190  402575/SN08000382  397513/SN08000185  407681/SN08000345  411751/SN08000438  410786/SN08000454  417251/SN08000547  427833/SN09000208  415998/SN08000511  416000/SN08000508  421979/SN09000098  422100/SN09000216  442036/SN09000336  448226/SN09000416  455130/SN10000286  462191/SN10000266  468071/SN10000398  467831/SN10000326  471707/SN11000023  477922/SN11000076  481951/SN11000205  489697/SN11000261",
      "center_classification_date": "20130617",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the country of Canada.",
      "state": "MA",
      "product_description": "Philips BuckyDiagnost Radiographic,  a Diagnostic X-ray system.    Catalog Numbers:    704031, 704032, 704035, 704060, 704062      This system is used for making x-ray exposures for diagnostics.",
      "report_date": "20130626",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Healthcare Inc.",
      "recall_number": "Z-1449-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64128",
      "termination_date": "20131015",
      "more_code_info": "",
      "recall_initiation_date": "20130116",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}