  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arras CEDEX",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83687",
      "recalling_firm": "Normand-Info S.A.S.U.",
      "address_1": "7 rue Frederic Degeorge",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: NY, AZ, FL, NJ, MA, OH, CA, IL, NC, OK, IN, KY, OR, MI, WI, PA, UT, GA, VA, TX, NE, SD, IA, MN, MD, DE, SC, MT, WV, LA, MO, TN, NM, NV, HI, WA, WY, AL, AR, MS, RI, ID, ME, DC, ND, AK, CO, CT, KS, NH, VT",
      "recall_number": "Z-1430-2020",
      "product_description": "Remisol Advance running on hardware with Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating systems",
      "product_quantity": "6479",
      "reason_for_recall": "A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.",
      "recall_initiation_date": "20190815",
      "center_classification_date": "20200303",
      "termination_date": "20210603",
      "report_date": "20200311",
      "code_info": "All serial numbers"
    }
  ]
}