  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Knittlingen",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91953",
      "recalling_firm": "Richard Wolf GmbH",
      "address_1": "Pforzheimer Str. 32",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of United Arab Emirates (AE), Argentina (AR), Austria (AT), Belgium (BE), BO, Brazil (BR), Canada (CA), Chile (CL), Czech Republic (CZ), Germany (DE), Denmark (DK), Ecuador (EC), Spain (ES), Finland (FI), France (FR), Greece (GR), Hong Kong (HK), Croatia (HR), Ireland (IE), India (IN), Italy (IT), Jordan (JO), Japan (JP), South Korea (KR), Kuwait (KW), Kazakhstan (KZ), LB, Lithuania (LT), Marzocco (MA), Mexico (MX), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Pakistan (PK), Poland (PL), Palestine (PS), Portugal (PT), Qatar (QA), Romania (RO), Serbia (RS), Russian Federation (RU), Saudi Arabia (SA), Slovenia (SI), Slovakia (SK), Thailand (TH), Turkey (TR).",
      "recall_number": "Z-1424-2023",
      "product_description": "FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 8735.685",
      "product_quantity": "319 units",
      "reason_for_recall": "Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.",
      "recall_initiation_date": "20230324",
      "center_classification_date": "20230424",
      "report_date": "20230503",
      "code_info": "UDI/DI: 04055207020265; Lot Numbers: 4500377618, 4500377296, 4500374239, 4500371819, 4500367415, 4500363102, 4500367414, 4500363666, 4500363102, 4500359744, 4500357831, 4500354319, 4500352415, 4500349678, 4500344619, 4500347121, 4500349678, 4500348605, 4500344619, 4500347121, 4500344619, 4500341300, 4500338554, 4500337773, 4500335110"
    }
  ]
}