  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90557",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide  Foreign:  Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam",
      "recall_number": "Z-1419-2022",
      "product_description": "ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant  Part Number: 254400525",
      "product_quantity": "872 units",
      "reason_for_recall": "Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring.  has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis,  Poor Joint Mechanics, Soft Tissue Irritation, Pain",
      "recall_initiation_date": "20220629",
      "center_classification_date": "20220720",
      "report_date": "20220727",
      "code_info": "GTIN: 10603295434061    Lots:  AB4536235, AB4733296,   AB4841493, AB4810808,   AB4841495, AB4904565,   AB4967214, AB5004557, and  AB5034303"
    }
  ]
}