  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.",
      "address_2": "",
      "product_quantity": "106",
      "code_info": "Serial No: 75070  750002  75046  750002  75114  75118  75127  75130  750021  75128  750015  75052  75039  75027  75067  75041  75028  75036  75034  75029  75031  75047  75048  75057  75054  75049  75056  75053  75058  75062  75064  75066  75068  76023  75037  76022  76024  75104  75111  75126  75112  750006  76008  750014  75059  750005  75061  75107  750007  750008  75129  750013  75120  750028  750016  750022  750024  750025  750026  750029  75115  75131  750023  7524  75119  76002  75121  75108  75102  75101  75108  75100  750004  75132  750019  750020  75063  75044  750001  75113  750012  75035  75038  75051  75051  750010  750017  75055  75060  75055  750027  75105  75042  75032  75069  75040  75116  75065  75050  75065  76011  76010  76016  76020  76021  750018",
      "center_classification_date": "20180417",
      "distribution_pattern": "Worldwide Distribution - U.S. Nationwide. International distribution to Canada, Mexico, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cuba, Czech Republic, Denmark, Djibouti, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Laos, Lebanon, Lithuania, Macedonia, Malaysia, Maldives, Mali, Martinique, Monaco, Mongolia, Myanmar, Nepal, Netherlands, New Calcedonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saint Pierre and Miquelon, Samoa, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.",
      "state": "OH",
      "product_description": "Brilliance Big Bore (Oncology) - Model no. 728243, Computed Tomography X-Ray system    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.",
      "report_date": "20180425",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-1410-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79666",
      "termination_date": "20200409",
      "more_code_info": "",
      "recall_initiation_date": "20180208",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}