  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Oranmore",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96398",
      "recalling_firm": "ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY",
      "address_1": "Deerpark Industrial Estate",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution nationwide.  International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.",
      "recall_number": "Z-1403-2025",
      "product_description": "Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627",
      "product_quantity": "90 US; 136 OUS",
      "reason_for_recall": "Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.",
      "recall_initiation_date": "20250303",
      "center_classification_date": "20250320",
      "report_date": "20250326",
      "code_info": "Item No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215;  Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219;  Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221;  Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809"
    }
  ]
}