  "results": [
    {
      "status": "Ongoing",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94083",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Rd",
      "address_2": "",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide distribution - US Nationwide and the countries of Canada, Chile, Colombia, Mexico, South Korea, Taiwan, Thailand.",
      "recall_number": "Z-1401-2024",
      "product_description": "Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A  Part Number:  6375  16010  62974  65246  66312  66342  66388  66650  66852  70472  74307  74365  74366  74430  74475  74477  74502  74773  74873  75065  75157  75195  75409  75410  75473  75533  75534  75607  75765  75797  75833  75903  75956  75973  76131  76178  76278  76306  76398  76427  76498  76519  76536  76565  76575  76628  76630  76635  76644  76687  76725  76775  76789  76832  76839  76878  76900  76999  77010  77014  77015  77105  77227  77299  77304  77305  77412  77450  77496  77516  77629  77722  77742  77757  77791  77794  77796  78018  78027  165720  205307A  65046-03  65080-06  65167-06  65934-01  65940-01  66041-01  66154-01  66503-02  70148-02  70178-06  70188-05  70638-07  70664-02  70665-01  70734-03  71064-01  71432-02  71433-01  71546-02  71733-02  71754-04  71801-02  71820-01  71928-06  71934-04  72073-02  72149-01  72226-06  72326-07  73008-02  73346-01  73567-01  73568-01  73776-04  73783-02  73882-03  74172-03  74685-01  74734-01  74833-01  74879-07  74921-02  74965-03  74965-04  74972-01  75030-01  75050-02  75068-01  75159-01  75293-01  75409-01  75409-02  75410-01  75410-02  75540-03  75540-05  75572-01  75642-02  75688-01  75790-02  75834-01  75927-01  75959-01  75972-01  76178-01  76273-01  76278-01  76308-01  76326-02  76372-01  76373-01  76374-01  76376-01  76491-01  76732-02  76873-02  77012-02  77127-01  77316-01  77496-01  77502-01  77521-01  77532-01  77655-01  77791-01  B75409  B75410  B75903  B76178  B76278",
      "product_quantity": "20,533 units",
      "reason_for_recall": "Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially  requiring surgical or medical/pharmacological intervention",
      "recall_initiation_date": "20240220",
      "center_classification_date": "20240329",
      "report_date": "20240410",
      "code_info": "Product Number/Product Name/GTIN/Lot Numbers: 16010                     CARDIOPLEGIA SET 4:1 RATIO NO RECIRC 00699753001534 2L05                           2L12                           2L19                           2P19                           165720;                    CARDIOPLEGIA SET 4:1 RATIO16 00699753000278 2K15                           2K22                           2K29                           2L05                           2L19                           2L26                           2M03                           2M10                           2P19                           3A02                           62974; CUSTOM CP50 4:1 CARDIOPLEGIA 00699753454880 0000321897 245106371                      302110840                      6375;                      CARDIOPLEGIA SET 4:1 RATIO WITH BRIDGE16 00699753008533 2K15                           2K22                           2L12                           2L19                           2L26                           2P19                           65046-03; PUMP AND TABLE PACK 00699753497092 0000311775 0000311776 243105334                      65080-06; ADULT-SSOCIAL-EL SALVADOR PAC 00699753517233 0000452207 65167-06; PUMP & TABLE FX PACK 00699753499348 0000329322 0000329331 0000447522 301104985                      303108168                      65246;                     CUSTOM 1:1 CONDUCER / MP4 00699753458147 2M17                           2P12                           65934-01; UCSD PTE CARDIOFLEX CP 1:4/4:1 00699753500808 0000288968 0000437413 0000452200 242102098                      243105189                      243105789                      246107180                      249109155                      251109777                      65940-01; UCSF ADULT PACK 00699753485297 0000328414 0000331083 243105281                      246105282                      250107937                      66041-01; VAMC SF TUBING PACK 00699753487833 0000329895 66154-01;                  PED CPG BT MUF LINE 00699753510548 303110004                      66312; 4:1 CARDIOPLEGIA SET 00699753489202 0000324255 242103295                      249107463                      308111832                      66342; RX15 NON X-COATED PACK 00699753491533 0000306497 0000306498 0000306499 248105327                      301108454                      302110899                      66388; RX25 NON X-COATED PACK 00699753493261 0000320596 0000329321 66503-02; ADULT PACK ROSALES FX25RE 00699753529069 0000450368 66650;                     CARDIOPLEGIA 00699753509344 251108577                      304110009                      304112395                      66852;                     CPG DEL NIDO PANAMA 00699753513716 243999665                      246105460                      251108579                      308112349                      70148-02; TOTAL VISION SYSTEM 00699753492615 0000309898 0000440663 302110835                      70178-06; AR CHILDRENS HYPO/INFANT SET 00699753510265 0000444141 0000457866 70188-05; UVRMC MEMBRANE PACK 00699753507319 0000306512 0000306513 0000441631 0000450286 250109165                      251107206                      302110344                      70472; COATED CUSTOM INFANT PACK 00699753461970 0000439896 70638-07; COMMUNITY OH 4:1 CARDIO CEN 00699753515789 0000324364 0000440665 0000457895 250107193                      251109771                      302110814                      303111435                      70664-02 ;                 PRESTIGE PACK 00699753507180 307114008                      307114219                      70665-01; PERFUSION PACK 00699753507197 0000321902 0000321903 305110370                      70734-03;                  X-COATED PRESTIGE PACK 00699753513815 245104987                      306109545                      306114105R                     71064-01; NEONATAL PACK 00699753508262 0000326891 0000326892 0000450287 71432-02; X-COATED FX NEONATE PACK 00699753501454 0000321878 0000324346 308114486                      71433-01; FX PEDIATRIC PACK WITH NO OXY 00699753501515 0000328612 71546-02;                VAMC-SAN ANTONIO 00699753490079 249107214                      71733-02; UT PERFUSION TUBING PACK 00699753484610 0000301072 0000301073 0000301074 0000303749 0000317609 0000317610 0000450374 0000450378 245104762                      248106367                      249106368                      251107869                      305112155                      71754-04; ANNS X-COATED FX PACK 00699753483248 0000453295 302105248                      71801-02; X-COATED ADULT PACK 00699753505964 0000307712 0000307713 0000307714 0000307715 0000440660 0000440675 0000450379 0000450380 0000453315 0000453316 248108420                      301109196                      302110321                      71820-01; VAMC PERFUSION TUBING PACK 00699753484627 0000303747 0000303748 245106538                      71928-06;                  CUSTOM X-COATED FX SMALL ADULT PACK 00699753512863 245104799                      71934-04; X-COATED PERFUSION PACK 00699753516168 0000441622 250108164                      305112783                      306111058                      72073-02; X-C PUMP & TABLE PACK WITH CP 00699753505797 0000326887 245103302                      250105485                      307111018                      72149-01; LAFAYETTE PERFUSION PACK 00699753479227 0000326890 0000331118 0000447525 248103303                      248105486                      303108169                      72226-06;                  CUSTOM X-C INFANT 1/4 X 1/4 BABY FX PACK 00699753512917 244104742                      72326-07; ROCKINGHAMOH W/1/2\"VENOUSCP 00699753516090 0000450283 243105199                      247108406                      249109151                      73008-02; X-COATED PUMP AND TABLE PACK 00699753516182 0000309060 73346-01; X-COATED PUMP AND TABLE PACK WITH FX25RW 00699753518032 0000440655 250108636                      73567-01; MEMORIAL HERMAN SMALL 00699753481664 0000453291 0000453331 73568-01;                  MEMORIAL HERMAN LARGE 00699753481657 304111470                      73776-04; X-COATED PRESTIGE PACK 00699753515598 0000329893 0000329894 250107924                      73783-02; X-COATED SMALL ADULT PACK 00699753516243 0000363564 0000450367 301110283                      73882-03; SALEMOHNO OXYCARDIO 00699753517431 0000450369 302108414                      303111422                      74172-03;                  X-COATED PERFUSION TUBING PACK 00699753516113 306112132                      306113046                      74307; SCRIPPS CEREBRAL PERFUSION PK 00699753486669 0000448562 74365; U OF M 3/8 X 3/8 ADULT PACK 00699753484696 0000309896 0000450384 302110850                      74366; U OF M 3/8 X 1/2 ADULT PACK 00699753484702 0000303782 0000303783 0000303784 0000307711 0000439909 245105292                      248105293                      248107923                      250109231                      303110884                      74430; X-COATED PUMP PACK W/DELPHIN 00699753489974 0000310922 0000310923 0000310924 0000321877 0000324345 245105849                      248106566                      251108832                      302110341                      304112138                      74475; X-COATED NEONATE PACK 00699753493520 0000326846 0000455782 243104974                      249107455                      301109528                      305111032                      74477; X-COATED ADULT PACK 00699753493544 0000448578 0000455771 74502; SANFORD XC FX PERFUSION PK 00699753487970 0000322412 250109166                      307112623                      74685-01;                  AIKEN-OH CARDIO 4:1 HEMOCONC 00699753515697 245106345                      74734-01;                  X-COATED FX25 HEART LUNG PACK 00699753490109 301104991                      304108170                      306111059                      74773;                     TERUMO LATIN AMERICA 00699753490314 246107286                      308112362                      74833-01;                  HCG-CARDIOPLEGIA 00699753517400 246105001                      74873;                    SANFORD XC SMALL PERF PACK 00699753490178 249107928                      307112143                      74879-07; HARRISBURG-1:4 PLEGIA XC 00699753517509 0000444101 0000453304 244106346                      250108828                      251109770                      302110809                      305112573                      74921-02; ST LUKE'S HOSPITAL 00699753499188 0000453299 250105289                      304105810                      74965-03; X-COATED HEART PACK 00699753499966 0000300333 0000300334 0000300335 247105303                      247107932                      249107219                      74965-04; X-COATED FX HEART PACK 00699753529076 0000440668 0000459155 74972-01;                  XC P&T PRESTIGE PK W FX25RE 00699753510098 249107225                      75030-01; X-COATED FX PACK 00699753503076 0000322984 75050-02;                  X-COATED PERFUSION PK W 1:4 CP 00699753512405 246106397                      75065; XC PRESCRIPTIVE PERF PACK 00699753493858 0000303785 0000303786 0000303787 0000309055 0000309056 245104840                      248107265                      251107945                      304110895                      75068-01; RWJ X-COATED PRES OXY PACK 00699753503021 0000324347 0000330691 0000447529 0000453301 0000459228 244104822                      247105313                      250107943                      75157; MORRISTOWN 1:4 CP50 CARDIOPLEG 00699753493728 0000326833 0000326870 0000440666 0000441573 246108411                      248109144                      251109765                      305112577                      75159-01; X-COATED RIO PERFUSION PACK 00699753499485 0000322394 306112153                      75195; CENTRAL MAINE 00699753494640 0000326872 246107229                      75293-01;                  1:4 CARDIOPLEGIA PACK 00699753515185 244105866                      75409;                    X-COATED FX15RW40 PACK 00699753497061 243101697                      75409-01;                  X-COATED FX15RW40 PACK         00699753516137 244104993                      75409-02;                  X-COATED FX15RW40 PACK 00699753517714 303109546                      307112346                      308110576                      75410;                     X-COATED FX25RW PACK 00699753497085 243101699                      75410-01;                  X-COATED FX25RW PACK 00699753516120 245104994                      307114541R                     75410-02;                  X-COATED FX25RW PACK 00699753517721 303109547                      307112347                      308110577                      75473; DSA/CP FX PACK 00699753497740 0000326884 251103305                      302105488                      305111065                      75533; MOTT NEONATE DEL NIDEO PACK 00699753498990 0000440664 75534; MOTT INFANT DEL NIDO PACK 00699753498969 0000453294 75540-03 ;                 WASHINGTON HOSPFX OXY CARDI 00699753515024 243105195                      246107185                      75540-05; WASHINGTON HOSP FX OXY CARD 00699753528741 0000441561 75572-01; X-COATED PERSCRIPTIVE FX PACK 00699753504516 0000331079 0000445810 245100846                      245102576                      75607; X-COATED PERFUSION PACK 00699753499980 0000330335 0000450310 75642-02; HEME X-COATED TUBING PACK 00699753504110 0000289021 0000289024 0000324348 0000324349 246106384                      304110829                      75688-01; X-COATED FX PRESCRIPTIVE PACK 00699753501782 0000457906 75765;                     X-COATED PEDIATRIC PACK 00699753507593 301108396                      75790-02; X-COATED FX PACK WITH CP50 00699753517424 0000439904 302110334                      304111757                      75797; X-COATED PERFUSION PACK 00699753529359 0000450281 0000455794 75833;                     X-COATED CPG 4:1 AND 1:4 PACK 00699753500273 302110002                      306112369                      309114431                      75834-01; MEDSTAR FX25 PUM;K AND 00699753516007 0000311783 0000311784 0000311785 0000441645 0000459159 242104755                      243104756                      245104757                      247104758                      249108447                      249109194                      304110881                      306112584                      306114124R                     306114374R                     307112649                      307114674R                     307114907R                     75903;                     X-COATED TRINIDAD FX25 PACK 00699753505209 249106074                      249106075                      75927-01; X-COATED MAIN; PACK 00699753505940 0000311778 0000311779 0000450299 0000453302 250107942                      251107245                      75956; CONDUCER/MP4 8:1 FOR FX OXY 00699753501713 0000453279 243101157                      243102725                      244103328                      249103341                      249105017                      249106083                      249106938                      301108194                      301108989                      301109523                      304110052                      304112721                      75959-01; X-COATED HARLINGEN BROWNSVILLE 00699753516229 0000311058 0000311651 247106394                      249107927                      75972-01; X-COATED FX LOW PRIME PACK 00699753515758 0000440678 0000459133 244105238                      244105239                      245106542                      247107266                      250109189                      250109190                      250109220                      250109221                      306114101R                     306114116R                     306114375R                     306114383R                     307113458                      307114537R                     75973; ALLEGIANCE FX PACK 00699753502895 0000324368 0000324369 302106395                      76131; OPEN HEART3/8X1/2CP50X-COAT 00699753504813 0000326843 0000328593 0000452204 0000452205 244105783                      247107192                      249108821                      249109153                      251109769                      301110303                      303111443                      305112578                      76178;                     ADULT BASIC PACK DOMINICAN REP 00699753506657; 251108129                      251108129A                     302108135                      302108135A                     302108136                      302108136A                     76178-01;                  X-COATED ADULT BASIC PACK DOMINICAN REP. 00699753517943 305108959                      305108961                      305108962                      305108963                      305108964                      306108960                      306108965                      76273-01;                  X-COATED FX25RW PACK 00699753507371 249104743                      76278;                     X-COATED FX1540 MASTER PACK 00699753504387 303108139                      303108139A                     76278-01;                  X-COATED FX1540 PERFUSION PACK 00699753521957 303111005                      303111006                      76306; X-COATED VICTORIA FX PACK 00699753505179 0000301082 76308-01; X-COATED FX PUMP & TABLE PACK 00699753517103 0000440647 0000441552 248106055                      251107470                      302108585                      303109527                      305110068                      306111028                      76326-02; X-COATED FX ADULT PACK 00699753508965 0000326852 0000326853 0000455789 76372-01; X-COATED NEONATAL FX PACK 00699753516205 0000455780 249106040                      249106059                      249107466                      301109531                      305111798                      307111029                      76373-01; X-COATED INFANT FX PACK 00699753516144 0000441524 301107467                      303110060                      303111030                      76374-01;                 X-COATED 3/8 PEDIATRIC FX PACK 00699753516038 244104978                      303108589                      307114783                      76376-01; XC CONGENITAL ADULT FX PACK 00699753517219 0000455766 307111031                      76398                     DEL NIDO P;K 00699753506602 244104339                      303110031                      306111095                      76427;                    X-COATED PUMP AND TABLE PACK 00699753507456 245105252                      301109206                      306114132R                     76491-01; X-COATED FX CENTRIFUGAL PUMP PACK 00699753513280 0000439891 0000459156 249106379                      307112154                      76498; X-C ST CLAIR FX TUBING PACK 00699753508705 0000326849 245105275                      250109212                      307114543R                     76519; X-COATED LAKELAND VSD FX PACK 00699753506091 0000297734 248108416                      250109157                      303111427                      76536 XC HAWAII FX25 LG ADULT PACK 00699753506497 0000322413; 0000441572 0000441659 248105190                      248106336                      248107187                      248107903                      249109147                      251109780                      302110310                      305112576                      76565; XC HAWAII FX15 SM ADULT PACK 00699753508149 0000326885 0000452194 250105191                      251109782                      305112579                      76575; MERCY X-COATED CONDUCER 00699753506855 0000433587 0000448598 76628                     X-CAOTED PUMP PACK 00699753511057 244106372                      247108434                      76630; HHL/LOURDES 00699753507036 0000324372 250104995                      306108175                      76635 ;                    X-COATED ADULT PACK 00699753507234 245106383                      250109217                      307114535R                     7664; X-COATED MIXED PACK 00699753507449 0000326582 76687; X-C LOW PRIME PACK W/CARDIO 00699753507609 0000326850 0000457882 249105255                      303110831                      76725;                     ECUADOR X-COATED CPG SET 4:1 AND 1:4 00699753509924 303110010                      76732-02;                  X-C FX ADULT PUMP & TABLE PACK 00699753512856 250998421                      251104772                      76775; ESSENTIA HEALTH FARGO 00699753508323 0000324377 0000452197 76789; X-COATED PUMP PACK 00699753508279 0000324337 250107916                      76832; X-COATED TUBING PACK 00699753510401 0000321900 0000321901 0000324373 0000324374 248106354                      248107920                      250106355                      306111530                      76839; XC RIO GRANDE VALLEY PERFUSION 00699753512283 0000450316 302109789                      76873-02; X-COATED BAYLOR ALL SAINTS 00699753509788 0000306501 0000453300 0000455803 0000459231 302110340                      76878; X-COATED RX PERFUSION PACK 00699753509535 0000450318 247105853                      76900; X-COATED PUMP PACK 00699753512481 0000303761 246106380                      76999;                     X-COATED HEART PACK 00699753511774 249104996                      77010; X-COATED OPEN SYSTEM FX CUSTOM 00699753510371 0000293006 0000293007 0000293008 0000293009 0000293010 0000293011 0000293012 77012-02;                  X-COATED CITIZENS VICTORIA FX 00699753514744 244106387                      250108392                      77014;                     X-COATED PERFUSION PACK 00699753513044 251107217                      77015;                     XC FX15 HEART AND LUNG PACK 00699753510111 249102139                      77105; X-COATED ADULT PACK 00699753511248 0000321904 0000321905 0000445824 245105200                      245105798                      248107173                      250109143                      301109192                      301110311                      302110818                      304111415                      307114533R                     307114534R                     77127-01; X-COATED ADULT FX PACK 00699753516076 0000440648 250103289                      250105446                      303108969                      304108583                      306109542                      307112421                      308111818                      77227; CPG-FVL 0000444114 307112389                      309114447                      77299; X-COATED 1/2X3/8 PUMP PACK 00699753512351 0000300327 0000328415 0000450387 244105267                      77304; X COATED TUBING PACK 00699753512870 0000326578 0000326579 0000453297 246105203                      308113637                      77305; XCOATED FX PRESTIGE PACK 00699753513389 0000444094 250103755                      302110825                      77316-01;                  X-COATED FPS/PRESCRIPTIVE PACK 00699753517998 248108206                      77412; X-COATED SEPARATE LINE ADULT 00699753513709 0000331082 0000441553 0000441650 0000455772 77450; X-COATED LARGE PATIENT PACK 00699753513655 0000441618 77496;                     X-COATED LEGACY PACK 00699753514669 250109264                      77496-01; X-COATED LEGACY PACK 00699753517301 0000440670 248105792                      250108826                      302110808                      77502-01; X-COATED TUBING PACK 00699753414716 0000440677 251107221                      77516; BON SECOURS RICHMONDOHOXYCP 00699753514102 0000309054 0000322389 0000322390 244105784                      247108415                      251109774                      302110813                      303111441                      77521-01;                 X-COATED AMC PACK 00699753517349 244105794                      250108827                      251109215                      301110301                      304110810                      307114691R                     77532-01; X-COATED ADULT FX PACK 00699753515994 0000455773 246105443                      302110065                      307111777                      77629 ;                    ADULT PRESTIGE PACK GUATEMALA 00699753515147 251105466                      77655-01; X-COATED HEAT LUNG FX PACK 00699753517530 0000440643 246102687                      249109193                      77722;                     X-COATED ST VINCENT DEL NIDO 00699753527492 308112342                      77742; X-COATED PERFUSION PACK WITH FX25RE 00699753517202 0000441619 301104998                      304108176                      77757;                     X-COATED UMBRELLA PACK 00699753515925 247107231                      306111756                      77791;                     X-COATED UMC ADULT FX25R HEART 00699753517394 250107911                      251108436                      77791-01; X-COATED UMC PERFUSION PACK 00699753527485 303111153                      77794 X-COATED OXY AND CP PACK 00699753518001 0000459143 304108563                      305113024                      77796 BAY MEDICAL-PANAMA CITY FL 00699753517448 0000444093 0000450351 78018 X-COATED ADULT PACK DOMINICAN REPUBLIC 00699753528482 0000450280 78027 ADULT SPEC PACK 00699753527881 0000444098 B75409                    X-COATED FX15RW40 PACK 00699753497054 243101697                      B75410                    X-COATED FX25RW PACK 00699753497078 243101699                      B75903                    X-COATED TRINIDAD FX25 PACK 00699753505216 249106074                      249106075                      B76178                    ADULT BASIC PACK DOMINICAN REP 00699753506664 251108129                      302108135                      302108136                      B76278                    X-COATED FX1540 MASTER PACK 00699753506664 303108139",
      "more_code_info": ""
    }
  ]
}