  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Hartland",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96344",
      "recalling_firm": "Merge Healthcare, Inc.",
      "address_1": "900 Walnut Ridge Dr",
      "address_2": "N/A",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US",
      "recall_number": "Z-1399-2025",
      "product_description": "VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.",
      "product_quantity": "11",
      "reason_for_recall": "When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.",
      "recall_initiation_date": "20250219",
      "center_classification_date": "20250320",
      "report_date": "20250326",
      "code_info": "Software Versions and Part Numbers:   10.0:  Part Numbers:  88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00,  88-00532-00, 88-00533-00, 70-00008-00;   10.1.1.1:  Part Numbers:  88-01094-00;   10.1.2:  Part Numbers:  88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00;   10.2.0:  Part Numbers:  88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00,  88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00,   10.2 P2:  Part Numbers:  88-01203-00  10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309"
    }
  ]
}