  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Charlotte",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82694",
      "recalling_firm": "KaVo Dental Technologies LLC",
      "address_1": "11727 Fruehauf Dr",
      "address_2": "N/A",
      "postal_code": "28273-6507",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide to AL, AR, AZ, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NV, OH, OK, PA, TN, TX, VA, WA, WI, WV and internationally to Canada.",
      "recall_number": "Z-1387-2019",
      "product_description": "DEXIS Titanium Intraoral Sensor    Product Usage:  The DEXIS Titanium Intraoral Sensor is a USB-driven sensor which is intended to acquire dental intraoral radiographic images. The DEXIS / KaVo sensor shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intra-oral radiographs. The DEXIS / KaVo sensor can be used either in combination with special positioning devices to facilitate positioning and alignment with the x-ray beam or it may also be positioned by hand with the assistance of the patient.",
      "product_quantity": "211 devices",
      "reason_for_recall": "The sensor's look up calibration files were incorrectly programmed during manufacturing, which can result in the reduction of image quality.",
      "recall_initiation_date": "20190429",
      "center_classification_date": "20190521",
      "termination_date": "20201125",
      "report_date": "20190529",
      "code_info": "Model Number CD63M115.CG; Part Number IOXS-1.012.6180; Serial Numbers: 13095594, 13095595, 13095599, 13095661, 13095665, 13095759,   13095762, 13095766, 13095769, 13095774, 13095780, 13095782, 13095783, 13095784, 13095786, 13095809, 13095810, 13095817, 13095823, 13095827, 13095839, 13095876, 13095890, 13095897, 13095899, 13095900, 13095921, 13095927, 13095929, 13095941, 13095942, 13095955, 13095957, 13096033, 13096034, 13096037, 13096038, 13096040, 13096063, 13096064, 13096065, 13096120, 13096128, 13096145, 13096151, 13096155, 13096162, 13096173, 13096193, 13096195, 13096200, 13096206, 13096212, 13096217, 13096218, 13096220, 13096223, 13096224, 13096227, 13096230, 13096234, 13096235, 13096244, 13096252, 13096257, 13096267, 13096272, 13096286, 13096290, 13096292, 13096295, 13096299, 13096311, 13096318, 13096322, 13096335, 13096340, 13096344, 13096346, 13096348, 13096355, 13096362, 13096383, 13096391, 13096402, 13096405, 13096410, 13096417, 13096426, 13096428, 13096430, 13096437, 13096439, 13096466, 13096479, 13096483, 13096490,   13096491, 13096504, 13096505, 13096518, 13096527, 13096564, 13096574, 13096576, 13096580, 13096583, 13096591, 13096593, 13096602, 13096610, 13096622, 13096725, 13096738, 13096760, 13096804, 13096825, 13096831, 13096883, 13096905, 13096916, 13096928, 13096947, 13096958, 13096960, 13096964, 13096968, 13096970, 13096975, 13096979, 13096982, 13096984, 13096993, 13097017, 13097018, 13097038, 13097039, 13097101, 13097109, 13097118, 13097121, 13097126, 13097153, 13097166, 13097170, 13097181, 13097190, 13097193, 13097195, 13097199, 13097211, 13097214, 13097217, 13097234, 13097240, 13097241, 13097247, 13097249, 13097252, 13097269, 13097287, 13097299, 13097306, 13097309, 13097320, 13097322, 13097326, 13097348, 13097386, 13097389, 13097398, 13097418, 13097419, 13097422, 13097438, 13097439, 13097464, 13097469, 13097476, 13097478, 13097503, 13097517, 13097518, 13097532, 13097546, 13097553, 13097554, 13097557,   13097559, 13097562, 13097580, 13097584, 13097589, 13097603, 13097615, 13097616, 13097619, 13097808, 13097827, 13097838, 13095874, 13095917,   13095930, 13096057, 13096240, 13096448, 13097145, 13097230, 13097245, 13097463, 13097813."
    }
  ]
}