  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87475",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland,   France, Germany, Hong Kong, Ireland, Italy, Netherlands,   Norway, Poland, Portugal, Spain, Sweden, Switzerland,   United Kingdom.",
      "recall_number": "Z-1385-2021",
      "product_description": "Covidien Signia Small Diameter Curved Tip Intelligent Reload  30 mm Vascular/Thin 8 mm - Short    Item Code:SIGSDS30CTVT",
      "product_quantity": "15298 units WW: 9204 US; 6094 OUS",
      "reason_for_recall": "Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.",
      "recall_initiation_date": "20210304",
      "center_classification_date": "20210409",
      "termination_date": "20240712",
      "report_date": "20210421",
      "code_info": "Lot Numbers:  N0A0274UY  N0D0037UY  N0G0491UY  N0K1016UY  N0A0897UY  N0D0391UY  N0G0736UY  N0L0077UY  N0B0640UY  N0E0175UY  N0J0510UY  N0M0063UY  N0B0937UY  N0E0802UY  N0J0680UY  N0M0259UY  N0C0144UY  N0F1055UY  N0J0999UY  N9K1094UY  N0C0486UY  N0G0185UY  N0K0584UY  N9L0795UY  N0C1115UY  N0G0250UY  N0K0805UY    UDI:  10884521741874  20884521741871"
    }
  ]
}