  "results": [
    {
      "status": "Ongoing",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91906",
      "recalling_firm": "Encore Medical, LP",
      "address_1": "9800 Metric Blvd",
      "address_2": "N/A",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: TX, IN, MD, CA, KS, AL, HI, WA",
      "recall_number": "Z-1384-2023",
      "product_description": "EMPOWR Porous Knee Patella Over Drill, REF: 802-05-104; EMPOWR Knee Patella Drill, REF: 802-05-108",
      "product_quantity": "29",
      "reason_for_recall": "Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.",
      "recall_initiation_date": "20230228",
      "center_classification_date": "20230412",
      "report_date": "20230419",
      "code_info": "REF/UDI-DI/Lot: 802-05-104/00190446238843/388620L01; 802-05-108/00190446365907/388620L03"
    }
  ]
}