  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87475",
      "recalling_firm": "Covidien, LP",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland,   France, Germany, Hong Kong, Ireland, Italy, Netherlands,   Norway, Poland, Portugal, Spain, Sweden, Switzerland,   United Kingdom.",
      "recall_number": "Z-1384-2021",
      "product_description": "Covidien Signia Grey Small Diameter Curved Tip Intelligent Reload 30 mm Vascular 8 mm - Short  Item Code: SIGSDS30CTV",
      "product_quantity": "11391 units WW: 5982 US 5409 OUS",
      "reason_for_recall": "Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.",
      "recall_initiation_date": "20210304",
      "center_classification_date": "20210409",
      "termination_date": "20240712",
      "report_date": "20210421",
      "code_info": "Lot Numbers:  N0A0191UY  N0D0392UY  N0F0406UY  N0L0708UY  N0B0657UY  N0E0133UY  N0G0907UY  N9K0376UY  N0C0349UY  N0E0956UY  N0J0746UY  N9K1093UY  N0C0492UY  N0E1026UY  N0K0248UY  N9L0796UY  N0C1008UY  N0F0202UY  N0K0803UY  N9M0418UY  N0D0283UY  N0F0221UY  N0L0307UY  N9M0651UY    UDI:  10884521741881  20884521741888"
    }
  ]
}