{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90314",
      "recalling_firm": "Ivenix, Inc.",
      "address_1": "50 High St Ste 50",
      "address_2": "",
      "postal_code": "01845-2620",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of NJ and WI.",
      "recall_number": "Z-1381-2022",
      "product_description": "Ivenix Infusion System (IIS)",
      "product_quantity": "1335 devices",
      "reason_for_recall": "Downstream occlusion alarm is sometimes immediately followed by a pump problem alarm.",
      "recall_initiation_date": "20220422",
      "center_classification_date": "20220713",
      "termination_date": "20240520",
      "report_date": "20220720",
      "code_info": "UDI-DI: LVP-SW-0004 Software version 5.0 System serial numbers: 2118200011 2119300411 2119300414 2119300415 2119300417 2119300418 2119300421 2119300422 2119300423 2119300424 2119300425 2119300426 2119300427 2119300428 2119300429 2119300430 2119300431 2119300432 2119300434 2119300435 2119300436 2119300437 2119300438 2119300439 2119300440 2119300441 2119300442 2124300050 2124300051 2124300052 2124300055 2124300058 2124300062 2124300067 2124300068 2124300070 2124300073 2124300074 2124300076 2124300078 2124300079 2124300081 2124300082 2124300084 2124300086 2124300087 2124300088 2124300089 2124300090 2124300092 2124300094 2124300095 2124300096 2124300101 2124300102 2124300105 2124300123 2124300124 2124300127 2124300133 2124300137 2124300141 2124300130 2124300114 2124300140 2124300115 2124300129 2124300100 2124300109 2124300110  2118200008 2119300412 2119300416 2124300049 2124300053 2124300054 2124300056 2124300059 2124300060 2124300061 2124300064 2124300065 2124300066 2124300069 2124300071 2124300072 2124300075 2124300077 2124300080 2124300083 2124300085 2124300091 2124300093 2124300097 2124300104 2124300106 2124300107 2124300108 2124300112 2124300116 2124300118 2124300119 2124300120 2124300121 2124300122 2124300125 2124300128 2124300131 2124300132 2124300134 2124300135 2124300138 2124300139 2124300142 2124300143 2124300144 2124300145 2124300146 2124300147 2124300148 2124300150 2124300152 2124300153 2124300154 2124300155 2124300156 2124300157 2124300159 2124300160 2124300161 2124300162 2124300163 2124300164 2124300166 2124300167 2124300168 2124300169 2124300170 2124300171 2124300172 2124300173 2124300174 2124300175 2124300176 2124300177 2124300178 2124300179 2124300180 2124300181 2124300183 2124300184 2124300185 2124300186 2124300187 2124300188 2125900425 2125900426 2125900427 2125900428 2125900429 2125900431 2125900432 2125900434 2125900435 2125900436 2125900437 2125900438 2125900439 2125900440 2125900441 2125900442 2125900443 2125900444 2125900445 2125900447 2125900448 2125900449 2125900450 2125900451 2125900452 2125900453 2125900454 2125900456 2125900458 2125900459 2125900460 2125900461 2125900462 2125900463 2125900464 2125900465 2125900466 2125900467 2125900468 2125900469 2125900472 2125900473 2125900474 2125900475 2125900476 2125900477 2125900478 2125900479 2125900480 2125900481 2125900483 2125900484 2125900485 2125900486 2125900487 2125900488 2125900489 2125900490 2125900491 2125900492 2125900493 2125900494 2125900496 2125900497 2125900498 2125900499 2125900500 2125900502 2125900503 2125900504 2125900505 2125900506 2125900507 2125900509 2125900510 2125900511 2125900513 2125900514 2125900515 2125900516 2125900518 2125900519 2125900521 2125900523 2125900524 2125900525 2125900526 2125900527 2125900528 2125900529 2125900531 2125900532 2125900533 2125900534 2125900535 2125900536 2125900537 2125900538 2125900539 2125900540 2125900541 2125900542 2125900543 2125900544 2125900545 2125900546 2125900547 2125900548 2125900549 2125900550 2125900551 2125900552 2125900554 2125900555 2125900556 2125900557 2125900558 2125900559 2125900560 2125900561 2125900562 2125900563 2125900564 2125900565 2125900566 2125900567 2125900568 2125900569 2125900570 2125900571 2125900573 2125900574 2125900575 2125900576 2125900577 2125900578 2125900579 2125900580 2125900582 2125900583 2125900584 2125900585 2125900586 2125900587 2125900588 2125900589 2125900590 2125900591 2125900592 2125900593 2125900594 2125900595 2125900596 2125900597 2125900598 2125900599 2125900600 2125900601 2125900602 2125900603 2125900605 2125900606 2125900607 2125900609 2125900610 2125900611 2125900612 2125900613 2125900614 2125900616 2125900617 2125900618 2125900619 2125900620 2125900621 2125900622 2125900623 2125900624 2130700826 2130700827 2130700828 2130700829 2130700830 2130700831 2130700834 2130700835 2130700836 2130700837 2130700838 2130700839 2130700840 2130700841 2130700843 2130700844 2130700845 2130700846 2130700847 2130700848 2130700849 2130700850 2130700851 2130700852 2130700853 2130700855 2130700856 2130700858 2130700859 2130700860 2130700861 2130700863 2130700864 2130700865 2130700866 2130700867 2130700869 2130700870 2130700871 2130700873 2130700874 2130700875 2130700876 2130700877 2130700878 2130700879 2130700880 2130700881 2130700882 2130700884 2130700885 2130700886 2130700887 2130700888 2130700889 2130700891 2130700893 2130700895 2130700897 2130700898 2130700900 2130700901 2130700902 2130700903 2130700905 2130700906 2130700908 2130700909 2130700910 2130700911 2130700912 2130700913 2130700914 2130700915 2130700916 2130700917 2130700918 2130700919 2130700920 2130700921 2130700922 2130700923 2130700924 2130700925 2130700926 2130700927 2130700928 2130700929 2130700930 2130700931 2130700932 2130700933 2130700934 2130700935 2130700936 2130700937 2130700938 2130700939 2130700940 2130700941 2130700942 2130700943 2130700944 2130700945 2130700946 2130700947 2130700948 2130700949 2130700950 2130700951 2130700952 2130700953 2130700954 2130700955 2130700956 2130700957 2130700958 2130700959 2130700960 2130700961 2130700962 2130700963 2130700964 2130700965 2130700966 2130700967 2130700969 2130700970 2130700971 2130700972 2130700973 2130700974 2130700975 2130700976 2130700977 2130700978 2130700979 2130700980 2130700982 2130700983 2130700984 2130700985 2135000505 2135000507 2135000511 2135000525 2135000543 2124300099 2124300103 2124300126 2201811326 2201811382 2201811406 2201811461 2201811463 2201811544 2201811587 2201811606 2201811607 2130700868 2130700894 2135000515 2135000527 2135000532 2135000536 2135000545 2135000517 2135000524 2135000538 2135000506 2130700981 2135000251  2135000518 2205900089 2205900093 2205000100 2205000119 2205000135 2205000137 2205000138 2205000140 2205000142 2205000144 2205000160 2205000165 2205000169 2205000173 2205000175 2125900581   2135000523 2135000526 2135000529 2135000530 2135000531 2135000533 2135000535 2135000540 2135000541 2135000542 2135000546 2135000549 2135000550 2135000553 2135000554 2135000560 2135000561 2135000563 2135000564 2135000566 2135000568 2135000569 2135000570 2135000571 2135000572 2135000575 2135000576 2135000577 2135000578 2135000581 2135000584 2135000585 2135000586 2135000588 2135000589 2135000590 2135000591 2135000593 2135000596 2135000598 2135000600 2135000601 2135000603 2135000604 2135000605 2135000606 2135000607 2135000609 2135000613 2135000615 2135000616 2135000618 2135000619 2135000620 2135000621 2135000622 2135000623 2135000624 2135000627 2135000628 2135000629 2135000630 2135000631 2135000632 2135000633 2135000634 2135000637 2135000641 2135000642 2135000643 2135000645 2135000647 2135000649 2135000652 2135000654 2135000655 2135000656 2135000658 2135000659 2135000660 2135000661 2135000663 2135000664 2135000667 2135000668 2135000669 2135000671 2135000672 2135000674 2135000675 2135000676 2135000677 2135000679 2135000681 2135000685 2135000687 2135000688 2135000689 2135000691 2135000692 2135000695 2135000696 2135000697 2135000698 2135000699 2135000700 2135000703 2135000704 2135000706 2135000707 2135000708 2135000710 2135000711 2135000712 2135000713 2135000714 2135000715 2135000716 2135000717 2135000718 2135000719 2135000720 2135000722 2135000724 2135000726 2135000728 2135000729 2135000730 2135000733 2135000735 2135000736 2135000737 2135000738 2135000739 2135000740 2135000742 2135000743 2135000747 2135000748 2135000749 2135000750 2136400556 2136400557 2136400559 2136400560 2136400562 2136400563 2136400564 2136400565 2136400566 2136400568 2136400569 2136400570 2136400571 2136400572 2136400573 2136400574 2136400576 2136400578 2136400580 2136400582 2136400583 2136400584 2136400586 2136400587 2136400588 2136400591 2136400593 2136400595 2136400596 2136400598 2136400600 2136400602 2136400606 2136400607 2136400610 2136400612 2136400613 2136400614 2136400616 2136400617 2136400618 2136400619 2136400620 2136400621 2136400622 2136400624 2136400625 2136400627 2136400628 2136400631 2136400633 2136400635 2136400636 2136400638 2136400639 2136400642 2136400643 2136400645 2136400646 2136400647 2136400653 2136400655 2136400656 2136400657 2136400658 2136400659 2136400660 2136400662 2136400663 2136400665 2136400667 2136400668 2136400669 2136400670 2136400673 2136400675 2136400676 2136400677 2136400679 2136400680 2136400681 2136400682 2136400686 2136400687 2136400689 2136400690 2136400691 2136400692 2136400695 2136400698 2136400700 2136400701 2136400704 2136400705 2136400706 2136400707 2136400708 2136400709 2136400710 2136400712 2136400713 2136400714 2136400716 2136400717 2136400718 2136400719 2136400720 2136400721 2136400727 2136400728 2136400730 2136400735 2136400737 2136400739 2136400741 2136400742 2136400743 2136400744 2136400746 2136400748 2136400749 2136400750 2136400751 2136400752 2136400754 2136400755 2136400757 2136400758 2136400760 2136400763 2136400764 2136400767 2136400769 2136400770 2136400772 2136400775 2136400778 2136400779 2136400780 2136400781 2136400782 2136400783 2136400784 2136400785 2136400791 2136400793 2136400794 2136400796 2136400797 2136400798 2136400799 2136400800 2136400801 2136400803 2136400804 2136400805 2201811308 2201811309 2201811310 2201811314 2201811316 2201811317 2201811319 2201811321 2201811327 2201811328 2201811331 2201811333 2201811336 2201811339 2201811340 2201811341 2201811342 2201811345 2201811348 2201811353 2201811354 2201811356 2201811359 2201811360 2201811361 2201811365 2201811367 2201811368 2201811369 2201811370 2201811371 2201811372 2201811375 2201811376 2201811378 2201811380 2201811381 2201811383 2201811384 2201811386 2201811387 2201811391 2201811392 2201811394 2201811396 2201811403 2201811404 2201811409 2201811410 2201811411 2201811415 2201811416 2201811419 2201811420 2201811425 2201811428 2201811429 2201811432 2201811433 2201811435 2201811444 2201811448 2201811451 2201811452 2201811454 2201811455 2201811459 2201811465 2201811467 2201811472 2201811481 2201811483 2201811485 2201811487 2201811497 2201811506 2201811507 2201811513 2201811515 2201811516 2201811518 2201811524 2201811534 2201811538 2201811539 2201811541 2201811554 2201811556 2201811557 2201811562 2201811564 2201811565 2201811570 2201811579 2201811585 2201811589 2201811591 2201811592 2201811596 2201811604 2201811605 2201811609 2201811618 2124300063 2125900553 2135000503 2135000508 2135000559 2135000573 2135000583 2135000597 2135000610 2135000625 2135000638 2135000639 2135000640 2135000644 2135000648 2135000650 2135000673 2135000701 2135000702 2135000705 2135000709 2135000727 2135000744 2136400589 2136400590 2136400605 2136400626 2136400640 2136400641 2136400652 2136400696 2136400734 2136400738 2201811320 2201811335 2201811346 2201811347 2201811349 2201811400 2201811401 2201811405 2201811414 2201811421 2201811422 2201811426 2201811427 2201811430 2201811439 2201811440 2201811442 2201811445 2201811473 2201811474 2201811480 2201811519 2201811523 2201811527 2201811529 2201811535 2201811540 2201811545 2201811550 2201811555 2201811560 2201811561 2201811573 2201811594 2201811598 2201811599 2201811601 2201811610 2201811613 2201811614 2201811616 2201811617 2201811621 2130700892 2135000501 2135000502 2135000504 2135000509 2135000510 2135000516 2135000520 2135000522 2135000528 2135000534 2135000548 2135000551 2135000552 2135000555 2135000556 2135000557 2135000558 2135000565 2135000567 2135000574 2135000579 2135000580 2135000587 2135000592 2135000595 2135000599 2135000608 2135000611 2135000612 2135000614 2135000617 2135000626 2135000636 2135000646 2135000651 2135000653 2135000657 2135000662 2135000665 2135000666 2135000670 2135000693 2135000680 2135000682 2135000684 2135000686 2135000690 2135000694 2135000721 2135000723 2135000725 2135000731 2135000732 2135000734 2135000741 2136400558 2136400561 2136400567 2136400575 2136400577 2136400579 2136400581 2136400585 2136400592 2136400594 2136400597 2136400599 2136400601 2136400603 2136400604 2136400608 2136400609 2136400611 2136400615 2136400629 2136400630 2136400632 2136400634 2136400637 2136400644 2136400648 2136400649 2136400650 2136400651 2136400654 2136400661 2136400664 2136400671 2136400674 2136400678 2136400683 2136400684 2136400685 2136400688 2136400693 2136400694 2136400697 2136400699 2136400702 2136400703 2136400711 2136400715 2136400722 2136400723 2136400724 2136400725 2136400726 2136400729 2136400732 2136400733 2136400736 2136400740 2136400747 2136400756 2136400759 2136400761 2136400762 2136400765 2136400766 2136400768 2136400771 2136400773 2136400774 2136400776 2136400777 2136400787 2136400788 2136400789 2136400790 2136400792 2136400795 2136400802 2201811307 2201811311 2201811312 2201811313 2201811318 2201811322 2201811323 2201811324 2201811325 2201811329 2201811330 2201811332 2201811334 2201811337 2201811338 2201811343 2201811344 2201811352 2201811357 2201811362 2201811363 2201811364 2201811366 2201811373 2201811379 2201811385 2201811388 2201811389 2201811390 2201811393 2201811395 2201811397 2201811398 2201811399 2201811402 2201811407 2201811412 2201811413 2201811417 2201811418 2201811423 2201811424 2201811431 2201811434 2201811436 2201811437 2201811438 2201811441 2201811443 2201811446 2201811447 2201811450 2201811453 2201811456 2201811458 2201811460 2201811462 2201811464 2201811466 2201811468 2201811469 2201811470 2201811471 2201811475 2201811476 2201811478 2201811479 2201811482 2201811484 2201811486 2201811489 2201811490 2201811491 2201811492 2201811493 2201811494 2201811495 2201811498 2201811499 2201811500 2201811501 2201811502 2201811503 2201811504 2201811505 2201811509 2201811510 2201811511 2201811512 2201811514 2201811517 2201811520 2201811521 2201811522 2201811525 2201811526 2201811528 2201811530 2201811531 2201811532 2201811533 2201811536 2201811537 2201811542 2201811546 2201811547 2201811548 2201811549 2201811551 2201811552 2201811553 2201811559 2201811563 2201811566 2201811567 2201811568 2201811569 2201811571 2201811572 2201811574 2201811577 2201811578 2201811580 2201811582 2201811583 2201811584 2201811586 2201811590 2201811593 2201811595 2201811597 2201811600 2201811602 2201811603 2201811608 2201811611 2201811612 2201811615 2201811619 2201811620 2205900060 2205900061 2205900062 2205900063 2205900064 2205900068 2205900073 2205900074 2205900077 2205900081 2205900082 2205900084 2205900086 2205900087 2205900090 2205900091 2205900092 2205900094 2205900096 2205900097 2205900098 2205900101 2205900102 2205900103 2205900104 2205900106 2205900108 2205900109 2205900110 2205900112 2205900113 2205900114 2205900115 2205900116 2205900117 2205900118 2205900120 2205900121 2205900122 2205900123 2205900124 2205900125 2205900126 2205900127 2205900128 2205900129 2205900131 2205900132 2205900133 2205900134 2205900147 2205900148 2205900153 2205900166 2205900172 2205900176 2205900067 2205900069 2205900072 2205900075 2205900076 2205900078 2205900083 2205900085",
      "more_code_info": ""
    }
  ]
}