  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87546",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "N/A",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Canada, Korea.",
      "recall_number": "Z-1377-2021",
      "product_description": "BD BBL Martin-Lewis Agar, 100 Ea - Product Usage: used for  the isolation of pathogenic Neisseria from specimens containing mixed flora of bacteria and fungi.  Catalog Number: 221558",
      "product_quantity": "15550 EA",
      "reason_for_recall": "Products show heavy growth of C. albicans after 24 hours of incubation due to misformulation (expected reaction is partial inhibition after 18-48 hours incubation)  and reduce the identification of Neisseria contributing to a delay in time to definitive diagnosis or a delay in proper treatment",
      "recall_initiation_date": "20210317",
      "center_classification_date": "20210408",
      "termination_date": "20220805",
      "report_date": "20210414",
      "code_info": "Lot Numbers:  0317559   Exp. Date 2021-MAR-11;   0345695   Exp. Date 2021-APR-06  UDI: 01)30382902215586 (10)0317559(17)031121 (30)0100  (01)30382902215586 (10)0345695(17)040621 (30)0100"
    }
  ]
}