  "results": [
    {
      "status": "Ongoing",
      "city": "Elkton",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91775",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "125 Blue Ball Rd",
      "address_2": "",
      "postal_code": "21921-5315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.  Belgium, Canada, Costa Rica, Thailand.",
      "recall_number": "Z-1345-2023",
      "product_description": "CAPIOX RX15 Hollow Fiber Oxygenator, Catalog Number 3CX*RX15RE30",
      "product_quantity": "1412 units",
      "reason_for_recall": "Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.",
      "recall_initiation_date": "20230301",
      "center_classification_date": "20230405",
      "report_date": "20230412",
      "code_info": "UDI-DI: (01)00699753450134; Lot Number: 2M10",
      "more_code_info": ""
    }
  ]
}