{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Queensbury",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94035",
      "recalling_firm": "Angiodynamics, Inc.",
      "address_1": "603 Queensbury Ave",
      "address_2": "",
      "postal_code": "12804-7619",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL,  NO, NZ, ZA.",
      "recall_number": "Z-1343-2024",
      "product_description": "MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75\"  PG-\tUsed for the percutaneous introduction of a guidewire into the vascular system.\t\t\t\t\t\t\t\t\t  \t\t\t\t\t  Catalog Number: 45-765",
      "product_quantity": "130 units",
      "reason_for_recall": "Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case",
      "recall_initiation_date": "20240130",
      "center_classification_date": "20240321",
      "report_date": "20240327",
      "code_info": "UDI: 15051684023085      UPN: H965457651     Lot Number:  5790681  5795961  5800479  5803518  5805067",
      "more_code_info": ""
    }
  ]
}