  "results": [
    {
      "status": "Ongoing",
      "city": "Wappingers Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98039",
      "recalling_firm": "Laerdal Medical Corporation",
      "address_1": "167 Myers Corners Rd",
      "address_2": "N/A",
      "postal_code": "12590-3869",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.",
      "recall_number": "Z-1341-2026",
      "product_description": "Laerdal Compact Suction Unit 4   RTCA version (LCSU 4   RTCA)  1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052;  2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062;  3) LCSU 4, RTCA; Catalog Number: 881152;",
      "product_quantity": "1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)",
      "reason_for_recall": "units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.",
      "recall_initiation_date": "20260113",
      "center_classification_date": "20260211",
      "report_date": "20260218",
      "code_info": "1) LCSU 4, 800 ml, RTCA;  Catalog Number: 880052; UDI-DI: 07045432088519;  2) LCSU 4, 300 ml, RTCA;  Catalog Number: 880062; UDI-DI: 07045432088533;  3) LCSU 4, RTCA;  Catalog Number: 881152; UDI-DI: 07045432067132;"
    }
  ]
}