{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90387",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr",
      "address_2": "N/A",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.",
      "recall_number": "Z-1341-2022",
      "product_description": "Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction.  Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C",
      "product_quantity": "113 units; Extended: 52 units",
      "reason_for_recall": "Positive bias with Lithium Heparin plasma samples from  individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction",
      "recall_initiation_date": "20220518",
      "center_classification_date": "20220706",
      "termination_date": "20250213",
      "report_date": "20220713",
      "code_info": "UDI-DI: 00842768013836GA232320221119  Lot Number: GA2323 Exp. Date: 11-19-2022  Recall Extended 9/21/22: UDI: 00842768013836BA308820230329  Lot Number: BA3088   Exp. Date: 2023-03-29",
      "more_code_info": ""
    }
  ]
}