{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90490",
      "recalling_firm": "Philips Medical Systems DMC GmbH",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24-26",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1329-2022",
      "product_description": "DigitalDiagnost C90 Flex/Value/Chest/ER.  radiography and fluoroscopy system",
      "product_quantity": "140 systems in total, (Updated 1/30/2023).",
      "reason_for_recall": "Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy.  (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.",
      "recall_initiation_date": "20210519",
      "center_classification_date": "20220712",
      "report_date": "20220720",
      "code_info": "Model 712035  Serial Numbers   21862105  21862107  21862123  21862126  22860027",
      "more_code_info": ""
    }
  ]
}