      "country": "United States",
      "city": "Bedford",
      "address_1": "35 Crosby Dr",
      "reason_for_recall": "The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.",
      "address_2": "",
      "product_quantity": "287",
      "code_info": "Serial numbers per recall strategy.",
      "center_classification_date": "20130516",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: DE, RU, AU, IT, GB, ES, FR, KR, ZA, CA, AR, LT, TW, CN, BE, AT, JP, QA, SA, SG, HK, SI, NL, and SE.",
      "state": "MA",
      "product_description": "Fluoroscan Mini C-arm InSight, software versions 1.x, 2.x, 3.x    The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including but no limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.",
      "report_date": "20130522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hologic, Inc.",
      "recall_number": "Z-1326-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65028",
      "termination_date": "20140821",
      "more_code_info": "",
      "recall_initiation_date": "20130425",
      "postal_code": "01730-1450",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}