{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90490",
      "recalling_firm": "Philips Medical Systems DMC GmbH",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24-26",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1325-2022",
      "product_description": "ProxiDiagnost N90.  radiography and fluoroscopy system",
      "product_quantity": "140 systems in total, (Updated 1/30/2023).",
      "reason_for_recall": "Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy.  (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.",
      "recall_initiation_date": "20210519",
      "center_classification_date": "20220712",
      "report_date": "20220720",
      "code_info": "Model 706100 - Serial Number - 10001002  10001003  10001009  10001012  10001014  10001017  10001019  10001020  10001021  10001023  10001027  10001028  10001030  10001037  10001038  10001039  10001040  10001041  10001042  10001043  10001044  10001047  10001049  10001051  20000079    Model 706110    10001052   10001058   10001062   10001068   10001087  10001057   10001059   10001064",
      "more_code_info": ""
    }
  ]
}