  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Yorba Linda",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84450",
      "recalling_firm": "Nobel Biocare Usa Llc",
      "address_1": "22715 Savi Ranch Pkwy",
      "address_2": "",
      "postal_code": "92887-4609",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.",
      "recall_number": "Z-1315-2020",
      "product_description": "Dense Bone Drill WP 5x13 mm Single-Pat - Product Usage: Drills are intended to be used in the upper and lower jaw bone to prepare osteotomy prior to implant placement.",
      "product_quantity": "None",
      "reason_for_recall": "Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.",
      "recall_initiation_date": "20190314",
      "center_classification_date": "20200219",
      "termination_date": "20240215",
      "report_date": "20200226",
      "code_info": "(01)07332747065459(10)73591   Catalog# 37098",
      "more_code_info": ""
    }
  ]
}