{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sparks",
      "address_1": "7 Loveton Cir",
      "reason_for_recall": "Background calibration has the potential to cause false positive or false negative test results. An incorrect test result with the BD Veritor System could lead to possible misdiagnosis and inappropriate treatment",
      "address_2": "",
      "product_quantity": "124",
      "code_info": "Serial Numbers:  Serial Number  1801179131J10  1801179119J10  1801179139J10  1801179117J10  1801179110J10  1801179115J10  1801159055J10  1801179120J10  1801159036J10  1801179109J10  1801179095J10  1801179128J10  1801189173J10  1801189181J10  1801189108J10  1801189146J10  1801189140J10  1801189126J10  1801189184J10  1801189159J10  1801179092J10  1801179137J10  1612139044GR0  1801179121J10  1801179124J10  1801179148J10  16102993246N0  1801159030J10  1801189102J10  1801189132J10  1801189166J10  1801169045J10  1801169012J10  1801169057J10  1801169049J10  1801169003J10  1801179155J10  1801179142J10  1710169021HM0  16102993166N0  1801169001J10  1801179096J10  1801179111J10  1801189176J10  1801189177J10  1801189180J10  1801169005J10  1801179094J10  1801179175J10  1801179113J10  1801179125J10  1801189104J10  1801189073J10  1801189082J10  1801169018J10  1801179157J10  1801179162J10  1801179156J10  1801189158J10  1801179105J10  1801179133J10  17122792607S0  1801169048J10  1801189183J10  1801189170J10  1801169051J10  1801169058J10  1801189127J10  1801179141J10  1801189178J10  1801189152J10  1801179144J10  1801179145J10  1603209035660  17121990337S0  1801179149J10  1801179165J10  1801159017J10  1801189106J10  1801179153J10  1801179154J10  1801179112J10  1801179101J10  1801189107J10  1801189168J10  16083090396J0  16083090446J0  16083090476J0  16083090766J0  16083190426J0  16083190656J0  1801189169J10  1801039026J00  1801189135J10  16083190756J0  1801169060J10  1801169056J10  1801169004J10  1801179130J10  1801179098J10  1801179114J10  1801179136J10  1801179134J10  1801179122J10  1703259291H60  1801189116J10  1604209189G80  1801159029J10  1801189063J10  1801189072J10  1801169019J10  1801179147J10  1801189182J10  1801189174J10  1801169050J10  1801169052J10  1801169006J10  1801189151J10  1801189061J10  1801189090J10  1801189167J10  1801189160J10  1801159039J10  1801159047J10",
      "center_classification_date": "20190510",
      "distribution_pattern": "Nationwide  Foreign:  Canada, Chile, Hong Kong, Japan, Taiwan",
      "state": "MD",
      "product_description": "BD Veritor System Reader -In Vitro Diagnostic for use with BD Veritor System Test Kit Devices  Catalog Number: 256055",
      "report_date": "20190522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Co.",
      "recall_number": "Z-1314-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82616",
      "termination_date": "20200504",
      "more_code_info": "",
      "recall_initiation_date": "20190403",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}