      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98232",
      "recalling_firm": "Fujirebio Diagnostics, Inc.",
      "address_1": "201 Great Valley Pkwy",
      "address_2": "N/A",
      "postal_code": "19355-1308",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.",
      "recall_number": "Z-1307-2026",
      "product_description": "Product Name: Lumipulse ¿-Amyloid Plasma Controls  Model/Catalog Number: 81559  Software Version: Not Applicable  Product Description: Lumipulse ¿-Amyloid Plasma Controls:  This quality control product can be used as an objective judgement of the laboratory s procedures and personnel techniques when analyzing ¿-amyloid1-42 in K2EDTA plasma on the LUMIPULSE G System. It is a valuable tool to assess good laboratory practices. Two levels of control are available to compare observations with expected ranges therefore assuring consistent performance of the testing system within the clinical range.    Lumipulse ¿-Amyloid Plasma Controls: Liquid (Frozen),  2 ¿2 Concentrations  L1      Level 1 (2¿ 1.5 mL)  L2      Level 2 (2¿ 1.5 mL)  Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative:  0.1% ProClin 950  Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio",
      "product_quantity": "179 units",
      "reason_for_recall": "Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).",
      "recall_initiation_date": "20251211",
      "center_classification_date": "20260205",
      "report_date": "20260211",
      "code_info": "Model/Catalog Number: 81559; UDI-DI: 05414470815593; Lot Number: T8B6116U;"
    }
  ]
}