  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94082",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "4100 Hamline Ave N",
      "address_2": "N/A",
      "postal_code": "55112-5700",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Austria, Belarus, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Netherlands, Poland, Portugal, Spain, and Japan",
      "recall_number": "Z-1302-2024",
      "product_description": "Boston Scientific POLARSHEATH Steerable Sheath 12F, REF M004CRBS3050",
      "product_quantity": "461 devices",
      "reason_for_recall": "Boston Scientific has identified a tooling error in manufacturing which may have caused delamination of the inner lumen of the sheath shaft in a subset of POLARSHEATH devices. This could result in the embolization of a fragment during use (such as flushing of the sheath or introduction of the dilator or ablation catheter).",
      "recall_initiation_date": "20240212",
      "center_classification_date": "20240326",
      "report_date": "20240403",
      "code_info": "GTIN 08714729992684, Batch Numbers:   32895945, 32895947, 32903274, 32903275, 32903276, 32903277, 32903278, 32903340, 32999267, 33000331, 33000333, 33000334,  33000337, 33000338, 33000339, 33052646, 33052647, 33052648, 33052649, 33061576, 33171976",
      "more_code_info": ""
    }
  ]
}