  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lexington",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98124",
      "recalling_firm": "FUJIFILM Healthcare Americas Corporation",
      "address_1": "81 Hartwell Ave Ste 300",
      "address_2": "N/A",
      "postal_code": "02421-3160",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.",
      "recall_number": "Z-1299-2026",
      "product_description": "FDR Visionary Suite;  Model Number: CH-200;  Version: (1) 566-16130-23, (2) 566-16130-31, (3)",
      "product_quantity": "16 units",
      "reason_for_recall": "It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.",
      "recall_initiation_date": "20260109",
      "center_classification_date": "20260205",
      "report_date": "20260211",
      "code_info": "Model Number: CH-200;  (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001;  (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001;  (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;"
    }
  ]
}