  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91652",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution list not yet provided.",
      "recall_number": "Z-1279-2023",
      "product_description": "Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178",
      "product_quantity": "10",
      "reason_for_recall": "There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.",
      "recall_initiation_date": "20230228",
      "center_classification_date": "20230322",
      "report_date": "20230329",
      "code_info": "Serial Numbers: MH353075E2 MH353057E2 MH353067E2 MH353077E2 MH356112G2 MH353056E2 MH356111G2 MH356119G2 MH353065E2 MH356117G2"
    }
  ]
}