  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90250",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution in states of Colorado, Florida, New Jersey, Indiana, Missouri, and Virginia",
      "recall_number": "Z-1279-2022",
      "product_description": "Flow-e Anesthesia System : Intended for use in administering anesthesia while controlling the entire ventilation of patient    Product Code/REF Number: 6887900",
      "product_quantity": "47 units US",
      "reason_for_recall": "Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.",
      "recall_initiation_date": "20220502",
      "center_classification_date": "20220623",
      "termination_date": "20240806",
      "report_date": "20220629",
      "code_info": "UDI-DI:  07325710010457  Serial Numbers: 50012, 50013, 50014, 50015, 50016, 50017, 50030, 50031, 50033, 50034, 50035, 50036, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50046, 50047, 50048, 50051, 50052, 50053, 50054, 50055, 50056, 50057, 50058, 50059, 50060, 50061, 50062, 50069, 50070, 50071, 50072, 50073, 50074, 50084, 50085, 50086, 50087, 50088, 50089, 50090, 50091, 50092, 50093, 50094, 50096, 50097, 50108, 50109, 50110, 50111, 50132. (Note: UDI was corrected July 7, 2022)",
      "more_code_info": ""
    }
  ]
}